daviteq A420-FCL User Manual Download Page 7

2.  DPD Total Chlorine Reagent (DPD Method #4) for measuring total chlorine.

The operator should review and familiarize themselves with the DPD method, reagents, and devices to make sure
knowing how to perform this important calibration measurement.
PreparetheequipmentandDPDreagentsforafastandefficientmeasurement.
MakesuretheA420-FCLchlorineelectrodeisfullypolarizedandconditionedinzero-chlorinewaterand  record  the  output
of zero point (the 1st point).
Make sure the electrode is installed properly inthe measurement system with a constant flowrate ranging from
30 L/hr to 60 L/hr (0.6 cm/s to 1.3 cm/s).
Add chlorine to the system, and preliminarily measure the chlorine concentration by DPDmethod.
Approximately adjust the chlorine concentration to the middle point of the measurement range by adding water
or chlorine if necessary.
Let  the  probe  output  be  stable,  and  then  take  water  samples  from  the  measurement  system  for  DPD
standardization. This sample should be close to the A420-FCL electrode location for best accuracy.
Record the output of A420-FCL (the 2nd point) as well as the chlorine reading from DPD test.
Calculate the slope as follows: 

Once the slope is obtained, the electrode output including voltage (0-5V) or current (4-20 mA) can be converted into
“mg/L” or “ppm” in customers’ analyzer/controller/meter. The conversion equation for any of A420-FCL output is
below:

When  the  zero  point  of  A420-FCL  is  not  within  specifications  or  there  is  no  significant  change  in  the  measurement
environment,  but  the  electrode  reading  is  continuously  increasing  or  decreasing  (not  fluctuating)  within  an  hour,  re-
polarization  is  probably  needed.  The  root  cause  is  that  some  interfering  substances  probably  were  brought  to  the
working electrode again, such as unsuccessfully polishing the surface of working electrode, changing the IFG, replacing
the membrane or soaking the electrode with the power off (unconnected) in chlorine or other strong oxidant solution
for a long time. The re- polarization process is mandatory until the electrode presents a stable output. If the zero point
of A420-FCL still can not meet the specifications after a reasonable amount of polarization time, changing the IFG or
polishing  the  working  electrode  again  may  be  necessary,  or  adding  tiny  sodium  thiosulfate  in  testing  water  for
removing the potential oxidizers.

If the zero point of the A420-FCL does not meet the listed specifications it may be necessary to change the IFG. The IFG
also needs be changed if the stainless-steel ring (counter electrode) is not submersed in the water when the electrode
is powered for more than 5 minutes. Also, when the slope is lower than 50% of the nominal value, changing the IFG
may  correct  this  situation.  It  is  important  to  disconnect/power  off  the  electrode  before  removing  it  from  the  water.
Instructions for changing the IFG are as follows:

Calibration Preparation and Procedure

5. Maintenance

5.1. Electrode re-polarization

5.2. Changing the Internal Filled Gel (IFG)

Summary of Contents for A420-FCL

Page 1: ...User Guide For Free Chlorine Sensor with Analog output A420 FCL User Guide For Free Chlorine Sensor with Analog output A420 FCL ...

Page 2: ...tration up to 50 ppm The bound membrane sensor cap provides good stability and durability for this product In addition our proprietary Internal Filled Gel IFG significantly reduces the electrode measurement s pH dependence In principle the chlorine electrode measures the total chlorine of the water sample via the measurement mechanism described below However free chlorine measurement is also feasib...

Page 3: ...le 6m with BNC connector Rating IP68 Sensor Dimension D21 3 x 219 mm Sensor net weight 200 grams Warranty 1 year for probe membrane for 03 months The Chlorine Electrode is designed to operate with 9 24 VDC power Connecting the electrode to any power outside the normal range may result in electrode damage and void the warranty In addition to complying with the instructions included in this manual b...

Page 4: ...aside Note Avoid damaging the Silver Halide Coating gray portion 4 Polish the working electrode WE surface see arrow with the polishing strip until it is shiny in appearance Polish the gold tip in one direction Do not over polish this should just take a few strokes on the polishing strip Avoid touching the silver halide coating Rinse the electrode with DI water after polishing 5 Fill the membrane ...

Page 5: ...The electrode is now ready for polarization and conditioning steps Yellow Vin Black Vin Red 4 20mA Input Green Ground 0VDC 1 The standard cable length is 3 feet If another cable length is needed contact your supplier or manufacturer The maximum length should be no more than 30 feet to avoid signal loss 2 The output current of power source should be more than 100 mA 3 The A420 FCL output can be mea...

Page 6: ...ration As a result a 2 point calibration determine the final calculation equation Usually the zero and the half scale points are recommended for calibrating A420 FCL For the zero point 1st point it can be obtained in zero chlorine water during the conditioning processing Generally the zero point of A420 FCL should be within the specification ranges provided in Tables 1 and 2 The zero point drift is ...

Page 7: ...ro point of A420 FCL is not within specifications or there is no significant change in the measurement environment but the electrode reading is continuously increasing or decreasing not fluctuating within an hour re polarization is probably needed The root cause is that some interfering substances probably were brought to the working electrode again such as unsuccessfully polishing the surface of wor...

Page 8: ...ane the membrane cap may need to be replaced Please make sure the electrode is not powered during assembly and then follow the instruction below for replacement 1 Unpack the membrane cap replacement kit and prepare the new membrane cap on a clean surface 2 Unscrew the old membrane cap and discard it 3 Rinse the new membrane cap with clean water then rinse it with 1 2 mL of the IFG 4 Follow the ins...

Page 9: ... outside of membranecap especially at center pore 3 Air bubbles inside the membrane cap 1 Clean or change membrane cap 2 Shake the electrode to remove theair bubbles from the center pore 3 Shake electrode or inspect the cap for air bubbles Slope too low 1 Membrane cap fouled 2 IFG deterioration 3 Insufficient conditioning 4 Flow rate too low 5 Erroneous DPD standardization 1 Clean or change membra...

Page 10: ... Hung Vuong Res An Lac Ward Binh Tan Dist Ho Chi Minh City Vietnam Tel 84 28 6268 2523 4 ext 122 Email info daviteq com www daviteq com Distributor in Australia and New Zealand Templogger Pty Ltd Tel 1800 LOGGER Email contact templogger net 8 Support contacts ...

Reviews: