David White Realist TR-300 4.5 Quick Start Manual Download Page 3

 

3

tangent screw, and then center the bubble by using the two leveling screws.  Rotate the 
telescope again 180° in azimuth.  The bubble should remain centered.  If it does not, 
repeat the process of bringing it halfway to center with the tangent screw and the rest of 
the way with the leveling screws.  When the bubble remains centered after the telescope 
is rotated, bring the telescope over the other pair of leveling screws and bring the bubble 
to center by using the leveling screws only.  This makes the vertical axis of the transit 
truly vertical. 

 

You may still have an error if the sighting is not parallel with the telescope bubble, or the 
vertical vernier is off.  You can detect this by making two sightings of the same distant 
feature with the telescope in normal position and reverted position (rotate the telescope 
vertically so the telescope bubble is on top) and comparing the readings.  Part of this can 
be fixed by adjusting the telescope bubble level, but the remaining 

index error

 (see 

below) is probably best resolved by making two readings, normal and reversed. 

 

6. Get the transit reading magnetic or true azimuths (though absolute accuracy is limited by the 
compass to about a degree – with transits and theodolites the absolute accuracy is not as 
important as the relative accuracy of angles) by: 

 

Using the upper clamp and tangent screw to set the direction to zero.  You'll need to 
understand how to read the horizontal circle vernier.  The lower scale shows the reading 
to a half degree; the upper shows the minutes above or below that reading.  For minutes 
find which mark lines up with a degree or half degree mark below.  The figure shows a 
reading of 33° 45'.   For now though, you just need to put the zero on the zero, but check 
the vernier to make sure the 0° mark is lined up with the 0' mark. 

 

 

 

 

With the upper clamp closed, and the compass released (unscrew its lock until it moves 
freely) open the lower clamp to rotate the telescope until the north arrow is pointing to 
north on the scale, or to the magnetic declination (may be around 15.5°) to get true 
directions.  Then close the lower clamp. 

 

Summary of Contents for Realist TR-300 4.5

Page 1: ...normal reading orientation Cross hair focus Compass Upper tangent screw Upper clamp Vertical clamp Focus Vertical tangent screw Lower clamp tangent screw Vertical circle vernier Plate levels Capstan screws top bottom Horizontal circle vernier Leveling screws Upper clamp tangent screw ...

Page 2: ...h the plate level you can level to greater accuracy with the telescope bubble Bring the telescope over a pair of leveling screws and center the telescope bubble by using the vertical clamp and tangent screw Rotate the telescope 180 in azimuth If the telescope bubble does not come to the center bring it halfway to the center by turning the vertical Adjusting a plate bubble level After centering the...

Page 3: ... bubble level but the remaining index error see below is probably best resolved by making two readings normal and reversed 6 Get the transit reading magnetic or true azimuths though absolute accuracy is limited by the compass to about a degree with transits and theodolites the absolute accuracy is not as important as the relative accuracy of angles by Using the upper clamp and tangent screw to set...

Page 4: ...fference is the index error which should remain reasonably consistent during a given use so you could also add this index error to the normal reading instead of making two readings but it s probably a good idea to check this on multiple readings Reading distances using the stadia rod If reading is at level the stadia readings can be converted to distances by subtracting the lower from the upper st...

Reviews: