OPERATING MANUAL

CALIBRATION CHECK

It’s possible for you to check the calibration of the laser. It’s recommended to check the 

calibration of the laser before performing any levelling task. Calibration by yourself is only 

approximate and the accuracy of the calibration will depend on the care with which you 

proceed. 

 

Place the instrument on a flat, level surface or a tripod 30m away from a wall or staff with a 

detector.

 

Align the first axis so that it’s square to the wall or staff. Allow the laser to self level 

completely.

 

Mark the position of the beam on wall or note the position of the beam from staff (P1).

 

Rotate the instrument 180° and allow the instrument to self level again. 

 

Mark or note the position of the beam (P2).

 

Rotate the instrument 90° and align the second axis so that it’s square to the wall and 

allow the laser to self level completely.

 

Mark the position of the beam on wall or note the position of the beam from staff (P3).

 

Rotate the instrument 180° and allow the instrument to self level again. 

 

Mark or note the position of the beam (P4).

 

The instrument is within its accuracy specification if the 4 parts are within ±1 mm from the 

centre. 

MAINTENANCE AND SERVICE

 

The instrument is a precision measuring device and should always be handled with care.

 

Do not attempt to open the device. To ensure product SAFETY and RELIABILITY, repairs, 

maintenance and adjustment should be performed by authorized service centres.  

Contact your dealer for more details. 

 

Do not drop, knock or shake the laser level as this will affect precision.

 

Remove batteries when the instrument is not in use. 

 

Keep the instrument clean at all times. Use a soft lint-free cloth for the body and lens tissue 

for glass windows. Do not use abrasive cleaners or solvents.

 

Do not store the instrument in the hot areas, hot temperatures can short the life of 

electronic device, damage battery, and warp or melt certain plastics.

 

Do not store the laser level in the cold areas. When the unit warms up to its normal 

operating temperature, moisture can form inside where it may damage circuit board.

 

Avoid shock, vibrations, extreme heat and humidity.

 

Do not immerse the instrument in water.

 

Always turn the instrument OFF when storing, transporting or moving around the job site.

 

Calibration is recommended at least once a year or as needed. 

 

When transporting the instrument, always make sure that it’s carried in its original tool 

case. Never carry the product loose in a road vehicle. 

Summary of Contents for DTR80R

Page 1: ...ROTARY LASER DIAL IN DUAL GRADE MODEL NO DTR80R DTR80G OPERATING INSTRUCTIONS TOPREVENT SERIOUS INJURY READ AND UNDERSTAND ALLWARNINGS AND INSTRUCTIONS BEFORE USE...

Page 2: ...CKLIST 5 COMPONENT OVERVIEW 6 LASER CONTROL DESCRIPTION 7 DETECTOR CONTROL DESCRIPTION 8 PRE OPERATION SET UP 9 OPERATION LASER INSTRUMENT 10 USING REMOTE FOR DIAL IN GRADE 12 OPERATION LASER DETECTOR...

Page 3: ...P65 TRIPOD THREADS 5 8 on side and base for horizontal and vertical operations REMOTE CONTROL DT100RM DETECTOR DT05REC DT06REC DATUM is constantly improving its products All information in this manual...

Page 4: ...r beam is aimed at a sturdy work piece without reflective surface Bright shiny reflective sheet steel or the like is not suitable for laser use as the reflective surface could direct the beam back at...

Page 5: ...y tool case with EVA foam 2 100m Remote control with LCD screen 3 Millimetre laser detector 4 Dual grade rotary laser 5 Ni MH battery charger indoor only 6 Alkaline battery pack 7 Staff mounting brack...

Page 6: ...eet for vertical operation without tripod 6 5 8 Tripod thread for vertical operation 7 Battery compartment 2 1 3 7 8 9 5 8 Battery charging jack for Ni MH battery only 9 Down plumb dot output window w...

Page 7: ...e from 0 10 45 90 180 To move the line press Rotation Control button A and B To exit the scan mode press the RPM button C E Tilt Alarm Wind Mode Press once to activate Tilt Alarm mode Press again to s...

Page 8: ...light Press and hold this button for 3 seconds to activate the backlight for extra visibility L Accuracy Setting Change detection accuracy LED INDICATOR DESCRIPTION ORANGE Indicates the laser beam is...

Page 9: ...he battery is full When the battery is running low all LED indicator lights on the control panel will start flashing in which the battery pack must be recharged WARNING Only use charger supplied by DA...

Page 10: ...or moved and the laser recommences spinning the instrument may not be at the original level position Please recheck your RL datum point RPM SPEED CONTROL Rotation speed can be controlled by pressing R...

Page 11: ...This function becomes active and monitors the movement of the laser 30 seconds after it s completely levelled and starts rotating This is used to prevent incorrect work or readings caused by movement...

Page 12: ...g Levelling icon on remote will start blinking while it s self levelling For other operation controls such as RPM speed control scanning mode and rotation control refer to previous section for details...

Page 13: ...are within 1 mm from the centre MAINTENANCE AND SERVICE The instrument is a precision measuring device and should always be handled with care Do not attempt to open the device To ensure product SAFET...

Page 14: ...impact damage Contact service centre for repair and calibration Instrument may be outside its self levelling range Adjust the position within the self levelling range Instrument may be in Manual Mode...

Page 15: ...ls They can be purchased separately and add more flexibility to get your job done HEAVY DUTY WALL BRACKET DT05AC DOME HEAD LASER TRIPOD DT1730D FLAT HEAD LASER TRIPOD DT1730F 5M E FACE STAFF DT5ME 5MM...

Page 16: ...product Consumer guarantees Our goods come with guarantees that cannot be excluded under the Australian Consumer Law You are entitled to a replacement or refund for a major failure and for compensati...

Reviews: