4 Operating Instructions
Setting the dial
w
w
w
w
WARNING
High percent dial settings combined with low gas flows may lead to
hypoxic mixtures in the breathing circuit. Datex-Ohmeda strongly
recommends the use of oxygen monitoring.
The dial release must be operated to turn the dial from the
setting.
Do not turn the dial if the vaporizer is not properly locked onto the
manifold.
Important
European Standard EN 740 -
Anesthetic Workstations and Their Modules
requires
that an appropriate gas monitor is used to monitor the concentration of anesthetic
agent vapor in the inspiratory gas when the vaporizer is in operation in order to provide
protection against hazardous output in the event of a device malfunction.
Datex-Ohmeda strongly recommends the use of anesthesia gas monitoring with this
equipment. Refer to local standards for mandatory monitoring.
OFF