Datex-Ohmeda S/5 monitors
Document No. 800 1009-1
44
5.2
Gases service menu
Noise Meas
A selection for activating the noise measurement.
Sample gain adj
A selection for adjusting the sampling pump gain, i.e. for for adjusting the sample flow
measurement.
Fall time Meas
A selection for activating the fall time measurement.
Pump ctrl
A manual control for the sampling pump
Zero valve ctrl
A manual control for the zero valve.
Occl valve ctrl
A manual control for the occlusion valve.
Record Data
Record Data prints out the shown service data and board information (id. serial number and
software id.) onto the recorder module XXX TAI MUUALLE, MIHIN? XXX, M-REC.
Service Data field
O
2
, CO
2
, N
2
O, AA
%
-field shows realtime concentrations,
noise-%
is standard deviation of concentration.
O
2
, CO
2
, N
2
O, AA channels A-E
mV
-field: signal is scaled to mV,
Gain
: User gain. It is scaled as (User gain)/(Factory gain).
ID
Shows the identified agent.
ID unrel.
The shown value tells how
unreliable
the identification is. With pure agent the value is normally <
50.
Sample Flow
Sample Flow
is calculated from differential pressure and is adjusted by the module.
Zero
value as
measured during initialization when the pump is off.
Gain
: sample flow measurement can be
calibrated by adjusting the gain.
Ambient
Ambient
: ambient pressure is measured every 30 min.
Amb-Work
: ambient pressure - sampling
system internal pressure.
Fall time
CO
2
and O
2
in ms. For N
2
O and AA same as CO
2
.