Datel muRata PS DMS-40LCD-4/20S Series Quick Start Manual Download Page 2

DMS-40LCD-4/20S Series

Ultra-Low 2.9V Loop-Drop 4½ Digit 

LCD Display Process Monitor

Technical enquiries

 email: [email protected], tel: 

+1 508 339 3000

www.murata-ps.com

MPM_DMS401LCD-420S.D01      

Page 2 of 5

TECHNICAL NOTES

DECIMAL POINT SETTINGS

Performance/Functional Specifications

Typical at T

A

 = +25°C,  unless otherwise noted.

(

1) Full Scale Input:

 The Display readings shown in Table 1 can  typically 

be obtained with transmitters having a low-level output of 3.8 to 4.3mA 
and a full-scale output of 19.4 to 20.4mA. When using a transmitter  
whose output falls outside these ranges and the desired display read-
ings are close to the upper or lower adjustment limits of the selected 
DIP-switch setting #, try using the next highest DIP-switch setting # 
if, after adjusting R7, the display reading is still too low, or the next 
lowest setting # if the display reading is still too high (see example 
number 2). Please keep in mind that the DMS-40LCD meter from 
which the DMS-40LCD-4/20S is derived has an accuracy specification 
of ±3 counts (max.); thus, it may not always be possible to obtain the 
exact desired display readings. A change of ±1 count is defined as the 
right-hand most digit going up or down by one. 

DMS-40LCD-4/20S   

4½ digit, LCD, loop-powered meter

 

DMS-BZL1

 

Panel-mount bezel assembly

DMS-BZL2

 

Panel-mount bezel assembly with 

 

sealing gasket

DMS-30-CP

 

Panel cutout punch

A panel-mount retaining clip is supplied with each meter.

Display readings other than those shown in Table 1 are obtainable. 
For example, some negative readings with a 4mA input are possible. 
Consult DATEL for more information regarding display readings not 
shown in the table.

(2) Max. Loop-Voltage Drop/Max. Loop impedance

: The maximum loop 

voltage-drop and maximum loop impedance (burden) are both speci-
fied with the meter configured for DIP-Switch setting #20 and an input 
loop current of 20.0mA.

(3) Temperature Drift:

 Temperature drift of zero and temperature drift of 

gain are both specified with the meter configured for range #20 with a 
4mA input adjusted (using R3) to read “0000” ±1 count, and a 20mA 
input adjusted (using R7) to read “19000” ±2 counts at an initial ambi-
ent temperature of 25°C.

The temperature drift of gain is proportional to the selected full-scale  
 range. It is typically less pronounced at the lower range settings, that  
 is, if the observed gain drift of a particular meter is +12 counts when  
 reading “19000” on range #20 at 40°C, the same meter will most  
 likely have a drift of only +6 counts when reading “9000” on range  
 #12 at 40°C. 

 The normally very accurate, temperature-insensitive autozero feature  
 of the DMS-40LCD-4/20S’s built-in analog-to-digital converter (A/D)  
 is not a significant factor in determining the meter’s overall zero- 
 reading stability over its rated operating temperature because an  
 offset voltage is applied to the A/D’s LO input. This offset is used to  
 null the signal voltage developed with a 4mA input. In order for the  
 A/D to display a steady “0000” ±1 digit, the autozero circuitry requires  
 both A/D inputs to always be at exactly zero volts. 

 The meter’s zero-reading stability over its specified operating  
 temperature is affected by the drift of three terms: the voltage    
 developed by the meter’s offset circuitry; the voltage developed  
 across the meter’s input resistors with a 4mA input; and the stability of  
 the applied 4mA input signal itself. In the lower DIP-switch settings  
 (ranges #1-4), the meter’s parasitic etch and switch impedances also  
 contribute a small error voltage. 

 The meter’s performance with regards to span stability over    
 temperature is affected by the drift of the meter’s gain circuitry and  
 loop input-resistors. To minimize these drifts in applications requiring  
 the utmost in temperature stability, where possible, the meter should  
 be calibrated at its anticipated operating temperature.

 Since the DMS-40LCD-4/20S uses extremely-stable thin-film chip  
 resistors, periodic re-calibration is typically required only in    
 environmentally demanding applications where shock, vibration,  
 and/or temperature extremes may have a detrimental effect on the  
 20-turn potentiometers. 

  Using the chart below as a guide, the DMS-40LCD-4/20S’s decimal  
  points can be configured to suit the user’s particular readout    
  requirements. To enable a desired decimal point place its DIP switch  
  to the ON position (up). Please note that the decimal points are 
  merely placeholders, that is, they can all be on or all off; they do not  
  affect the meter’s operation and/or display readings. 

Current Loop Input

Min.

Typ.

Max.

Units

Full Scale Input Range (1)

+3.8

+20.4

mA

Loop Burden (2)

190

Ohms

Voltage Drop (2)

2.9

3.8

Volts

Overcurrent Protection

±40

mA

Performance

Sampling Rate

2.5 reading per second

Accuracy (1 minute warm-up)

±0.05%FS ±1 Count

Temperature Drift of Span (3)

±0.3

±0.6

Cnts/°C

Temperature Drift of Zero (3)

±0.2

±0.5

Cnts/°C

Display

Display Type and Size

4½ digit, 0.4"/10.2mm high LCD

Polarity Indication

"–" for negative readings

Overrange Indication

"–1_ _ _" for negative inputs

"1_ _ _" for positive inputs

Physical/Environmental

Operating Temperature

0

+50

°C

Storage Temperature

–20

+75

°C

Humidity (non-condensing)

0

95

°C

Case Material

Polycarbonate

Weight

0.1 ounces (28 grams)

Ordering Information

Reviews: