5.7. Connecting to PC via USB cable
The following figure shows the wiring plan, connecting the instrument
to The PC:
1. Open the front cover of the instrument (from bottom to top).
2. To communicate with the configuration program, plug the included
USB cable (USB-B mini, 5-pin connector) into the "USB" socket
marked with the number (1) in the figure above. The design of the
cable is such that it can only be connected to the instrument in one
position.
The same applies to the cable end on the PC side.
3. Make the desired settings as described in chapter
7. Setting-up
.
USB driver installation is not required. It is not necessary to supply the
instrument with power while making the settings. After completing the
settings, simply unplug the USB cable from the instrument. (Don't
move it sideways or jerk it, just pull it.)
The USB port and the input of the instrument are not galvanically
isolated, so the PC can only be connected to the instrument under
operating conditions if the potential difference between them is
negligible (a balancing current of less than 1 mA flows between the
PC and the instrument).
It is not necessary to turn off the instrument when connecting or
disconnecting the USB cable.