DI-2008 Hardware Manual
Controls, Indicators, and Connections
15
Once the switch closure or TTL signal is connected, activate Remote Events through W
IN
D
AQ
Acquisition Software.
Events may be automatically placed on the rising or falling edge of the trigger signal. Use the menu command
Options
>
Remote
to set W
IN
D
AQ
to place event markers on low-to-high transitions of the Event input.
Use the menu command
Options
>
Remote Events -
to set W
IN
D
AQ
to place event markers on high-to-low transi-
tions of the Event input.
Event markers may also be placed in your data file manually by pressing the push button on the DI-2008 Instrument.
You must enable Remote Events in W
IN
D
AQ
to use the button (use the menu command
Options
>
Remote Events
+
or
Options
>
Remote Events -
).
An example event marker in a W
IN
D
AQ
data file is shown below.
*Please Note:
Event Markers do not display in the real-time W
IN
D
AQ
acquisition software - they only display in the
W
IN
D
AQ
playback software (WWB). Acquisition starts go up; Event markers spike downward.
www.dataq.com
USB
CH8
CH7
CH6
CH5
CH4
CH3
CH2
CH1
Download software at run.dataq.com
+15V GnD
60mA max
D0
D1
D2
D3
D4
D5
D6
DI-2008 Voltage, Thermocouple, and Pulse DAQ
Button
LED
0-30V Max
Count
Rate
Record
Event
120V rms max