2.4.2 CONTROL BOX CONNECTOR (LASER CONTROL)
Panel socket Sub-D, 15 positions, female.
Figure 11: Female panel socket Sub-D15 (front view).
PIN
SIGNAL
TYPE
(***)
DESCRIPTION
1, 2
5V DC
Output power
supply
Auxiliary 5V DC power supply available for drive input logical
HIGH (
max 200mA
)
3
SHUTTER EN
Digital Input
External “
Shutter
Enable
” signal (see paragraph 2.4.2.1):
- HIGH level: Shutter opened
- LOW level or disconnected: Shutter closed
4
START (*)
Digital Input
This signal is used to start to the marking process when a
document or a sequence is running in AUTO MODE (**) or
WORK MODE (**) (see paragraph 2.4.2.2):
- HIGH level pulsed signal start the marking process
5
STOP (*)
Digital Input
This signal is used to stop the marking process (see
paragraph 2.4.2.2):
- HIGH level pulsed signal stop the marking process
6
READY (*)
Digital
Output
This signal is used to know if a document, sequence or script
is loaded and ready to be executed:
- ON when a document or a sequence is running in AUTO
MODE (**) or WORK MODE (**) (SW_READY
COMPATIBILITY (**) = true)
- ON when a document or a sequence is running in AUTO
MODE (**) and laser in
READY
state (SW_READY
COMPATIBILITY (**) = false)
- ON when a script is running n AUTO MODE (**) and
“IoPort.setReady (true)” function is used
7
BUSY (*)
Digital
Output
This signal is used to know if the current spooler is executing
(marking in progress) (see paragraph 2.4.2.2):
- ON during marking process
8
END (*)
Digital
Output
This signal is used to know if the marking process is finished
(see paragraph 2.4.2.2):
- ON at the end of marking process
9
RESERVED
CAN_PWR
DO NOT CONNECT
10
KEY
Digital input
External “
KEY
” signal (see paragraph 2.4.2.1):
- HIGH level: system enabled;
- LOW level or disconnected: system disabled
11
RESERVED
CAN_H
DO NOT CONNECT
12
SHUTTER OPEN/
ALARM
Digital
Output
This signal is used to know if the system is ready to emit
laser radiation (see paragraph 2.4.2.1):
- ON when the system is in
READY
state
13
RESERVED
CAN_L
DO NOT CONNECT
14
RESERVED
CAN_GND
DO NOT CONNECT
15
GROUND
Ground
Ground
(*) refers to Lighter user’s manual “Setting I/O parameters” paragraph to set the signal properties
(**) refers to Lighter user’s manual
(***) refers to paragraph 2.5
NOTE:
Connection example in paragraph 2.6.