Datacolor Conditioner Operation Manual / DCCOPMN_RP1 / Rev 3.0 / 7th August 07
page 32 of 51
•
To unlock press the
and
buttons simultaneously for 3 seconds to move from the basic display
to the menu for selecting the Conditioner’s operation modes
•
Press button to enter the “operating mode hand”. In the upper display the value “OP.H” for
“operating mode hand” is displayed (Figure 7).
•
Press
button . The menu to key in the password is displayed (Figure 8).
Figure 8: Menu password
•
Key in the password by pressing
and
buttons.
•
Press
button
to confirm the change of the operating mode and to go back to the basic display (see
Figure 5). Changing of operating parameters according to chap. 6 is temporarily possible: 30 seconds
after the last keypad entry, the entry of operating parameters is locked again (operating mode “OP.L”).
If you want to permanently unlock the entry, set parameter “PA.H” to 0 (Table 3 on page 30 or Table 4
on page 31).
7.
Temperature safety devices
7.1
Electric safety controller (temperature safety device class 3.1)
The sample conditioning cabinet is equipped with an electric over temperature safety device class 3.1
acc. to DIN 12880. It is called “safety controller”. It serves to protect the charging material against
extensive high temperatures.
The safety controller is electrically independent from the main temperature controller and takes over
control when a selectable maximum temperature is reached. This maximum temperature is to be set at
the controller via the parameter “Al.5” (Table 3 on page 30 or Table 4 on page 31).
In case the safety controller is activated, an “absolute limit alarm” is triggered (chapter 8.1.1 on page 35).
On the upper controller display the message “otc” and the current interior temperature are displayed
alternately. Heating, humidification, and illumination (only CONDITIONER
TM
with lighting unit) are
automatically switched off and an alarm sound is emitted which can be muted by hitting controller button
.
When the interior temperature drops by 1°C below the value of the parameter “Al.5”, the absolute limit
alarm is terminated and the heating resets automatically. On the upper display the current interior
temperature is displayed only.
The necessary actions to be taken, if an absolute limit alarm occurs are described in chapter
8.1.1 on page 35)