11.1.2 Alarm messages
Condition
Alarm message
Start after
condition occurred
Zero-voltage relay
alarm output (option)
The current actual temperature value
is outside the tolerance range (chap.
11.4)
“Temperature range
after configurable
time
time as alarm start
The current actual humidity value is
outside the tolerance range (chap.
11.4)
“Humidity range
after configurable
time
time as alarm start
Cabinet door open
“Door open
after 5 minutes
----
Power failure
---
----
immediately
Exceeded setpoint of the safety control-
ler class 3.1
“Safety controller
immediately
----
Temperature sensor defective
e.g. “- - - - “ or
“<-<-< “ or “>->->”
immediately
----
Safety controller temperature sensor
defective
Safety controller
sensor
immediately
----
Alarm messages are displayed in the list of active alarms until acknowledging them. They are also shown
in the event list.
11.1.3 Messages concerning the humidity system
Condition and measures
Message
Start after condition
occurred
The humidity module is defective. Contact your local
Datacolor service representative.
“Humidity system”
immediately
The humidity module cannot fill up.
In case of freshwater supply via water pipe:
The water tap is closed, or the cabinet is defective (e.g.
inlet valve of humidity module).
In case of freshwater supply via freshwater can (option,
chap. 4.5):
Water can is empty. Humidification is turned off. In case
of refrigerating operation, the interior is strongly dehu-
midified. When the water supply is functional again, the
humidity system restarts, or the cabinet is defective.
“Freshwater supply”
immediately
The humidity module cannot empty the condensate
tank.
Wastewater tube obstructed. Check the length and loca-
tion of the wastewater tube. If appropriate contact your
local Datacolor service representative.
“Waste water”
immediately
Purging is required soon. Turn off and on again the
humidity switch to start purging. After a successful purg-
ing the notification resets automatically.
“Humidity system
purging req.”
after predefined time
(approx. 3-5 months
depending on use)
Maintenance of the humidity system is required. Contact
your local Datacolor service representative.
“Humidity module
service”
after predefined time
(approx. 1 year)
Messages concerning the humidity system are shown in the event list.