Chapter 6
100
Reference Triggers
Table 17: Reference Trigger Options
DT9837 Series
A/D
D/A
DIN
DOUT
C/T
TACH
QUAD
External Positive TTL Trigger Support
SupportsPosExternalTTLTrigger
External Negative TTL Trigger Support
SupportsNegExternalTTLTrigger
Positive Threshold Trigger Support
SupportsPosThresholdTrigger
Yes
a
a. If you choose a threshold trigger as a reference trigger on any of the DT9837 Series modules, you can program the threshold
level as a value between ±10 V for a gain of 1 or ±1 V for a gain of 10, and any of the supported analog input channels can be
used as the threshold channel.
Negative Threshold Trigger Support
SupportsNegThresholdTrigger
Yes
Digital Event Trigger Support
SupportsDigitalEventTrigger
Sync Bus Support
SupportsSyncBusTrigger
Analog Input Channels Supported for the Threshold Trigger
SupportedThresholdTriggerChannels
0 to 3
Post-Trigger Scan Count Support
SupportsPostTriggerScanCount
Yes
b
b. If you use a reference trigger, you can specify how many post-trigger samples to acquire after the reference trigger by
specifying the post-trigger scan count in software.
Summary of Contents for DT9837
Page 1: ...DT9837 Series UM 22417 AJ User s Manual Title Page ...
Page 4: ......
Page 12: ...About this Manual 12 ...
Page 26: ...Chapter 1 26 ...
Page 27: ...Part 1 Getting Started ...
Page 28: ......
Page 53: ...Verifying the Operation of a Module 53 ...
Page 60: ...Chapter 4 60 ...
Page 61: ...Part 2 Using Your Module ...
Page 62: ......
Page 90: ...Chapter 5 90 ...
Page 104: ...Chapter 6 104 ...
Page 110: ...Chapter 7 110 ...
Page 130: ...Appendix A 130 ...
Page 144: ...Appendix C 144 ...
Page 150: ...Index 150 ...