Specifications
99
Voltage Measurement Specifications
lists the voltage measurement specifications for the analog input subsystem on the
DT9828 module.
Table 26: Voltage Measurement Specifications
Feature
Specifications
a
a.
Conditions for specification:
Module is calibrated at 23 C.
Values are typical for the module operated in a stable 23°C ±5°C environment.
Valid for 1 year from calibration.
Number of voltage channels
Up to 8 differential voltage inputs
A/D converter resolution
24-bits
Full-scale input range
±156 mV
Aggregate sample rate
600 Samples/s maximum
Sampling mode
Multiplexed
DC differential input impedance
10 M
Ω
Input bias current
250 nA
Input common mode voltage
±500 V
Differential input voltage
±10 V maximum (protection)
Common mode rejection @ 60 Hz and 50 Hz
120 dB
Noise
3
μ
V pk-pk at maximum throughput (see
for more information)
Input signal bandwidth, –3 dB
Total throughput rate x 2, in Hz (see
for more information)
Offset error
b
b. Voltage source impedance <= 50 ohms
.
±20
μ
V
Gain error
±0.03% of reading
Linearity error
±2
μ
V
Offset error tempco
±20 nV/°C
Gain error tempco
±0.0015% of reading/°C
Summary of Contents for DT9828
Page 1: ...DT9828 User s Manual UM 24984 A Title Page ...
Page 4: ......
Page 8: ...Contents 8 ...
Page 12: ...About this Manual 12 ...
Page 13: ...13 1 Overview Features 14 Supported Software 16 Accessories 18 Getting Started Procedure 19 ...
Page 20: ...Chapter 1 20 ...
Page 21: ...Part 1 Getting Started ...
Page 22: ......
Page 30: ...Chapter 2 30 ...
Page 49: ...Verifying the Operation of a Module 49 ...
Page 54: ...Chapter 4 54 ...
Page 55: ...Part 2 Using Your Module ...
Page 56: ......
Page 57: ...57 5 Principles of Operation Analog Input Features 59 Digital I O Features 67 ...
Page 80: ...Chapter 6 80 ...
Page 86: ...Chapter 7 86 ...
Page 92: ...Chapter 8 92 ...
Page 106: ...Appendix A 106 ...
Page 112: ...Appendix B 112 ...