Calibrating the Analog Output Subsystem
This section describes how to use the DT9812 Series Calibration Utility to calibrate the analog
output subsystem of a DT9812-10V, DT9812-10V-OEM, DT9812A, DT9813-10V, DT9813A,
DT9814-10V, orDT9814A module.
To calibrate the analog output circuitry, you need to connect an external precision voltmeter to
analog output channels 0 and 1 of the module.
Do the following to calibrate the analog output circuitry:
1.
Select the
D/A Configuration
tab of the DT9812 Series Calibration Utility.
2.
Connect an external precision voltmeter to Analog Output 0 (DAC Ch0) of the module.
3.
Adjust the offset by entering values between 0 and 63 in the
DAC 0 Offset
edit box, or by
clicking the up/down buttons until the voltmeter reads 0 V.
4.
Connect an external precision voltmeter to Analog Output 1 (DAC Ch1) of the module.
5.
Adjust the offset by entering values between 0 and 63 in the
DAC 1 Offset
edit box, or by
clicking the up/down buttons until the voltmeter reads 0 V.
Note:
At any time, you can click
Restore Factory Settings
to reset the D/A calibration
values to their original factory settings. This process will undo any D/A calibration settings.
Once you have finished this procedure, the analog output circuitry is calibrated. To close the
DT9812 Series Calibration Utility, click the close box in the upper right corner of the window.
Summary of Contents for DT9812 Series
Page 1: ...DT9812 UM 20769 R DT9813 and User s Manual DT9814 Series Title Page ...
Page 4: ......
Page 12: ...About this Manual 12 ...
Page 13: ...13 1 Overview Key Hardware Features 14 Supported Software 15 Getting Started Procedure 16 ...
Page 17: ...Part 1 Getting Started ...
Page 18: ......
Page 26: ...Chapter 2 26 ...
Page 53: ...Part 2 Using Your Module ...
Page 54: ......
Page 76: ...Chapter 5 76 ...
Page 88: ...Chapter 6 88 ...
Page 94: ...Chapter 7 94 ...
Page 113: ...113 B Screw Terminal and Connector Pin Assignments ...