4.4 Fuse Replacement
CAUTION: Fuses normally blow for a reason. Investigate likely faults before changing fuses.
CAUTION: Isolate the unit from the mains before accessing the fuses.
Both the live and neutral mains inputs are fused. The fuses are contained in a holder which is
part of the IEC inlet mounted on the rear panel of the mobile base.
Unplug and remove IEC cable. Using a flat-bladed screwdriver, or similar, pull forward the
‘drawer’ containing the fuses. Hinge down the front of the holder. Lift the fuses by pushing
them through the holes below.
Refit new fuses (D10873) and reverse the removal sequence, ensuring the ‘drawer’ is
pushed fully home.
Warning: Ensure you isolate the unit from the mains power before any
work related to the electrical supply.
'Drawer'