Safety precautions

WARNING!
Due to safety considerations the 
needle (5) + (12) should always be 

placed in the retainer on the CheckPoint unit 
after measuring.
Use only the original external power supply 
unit / charger when connecting to mains.
Using the wrong type can destroy the 
CheckPoint unit. Use only the specified 
battery type or an original rechargeable 
battery pack. 
If rechargeable batteries have been installed 
in the CheckPoint, they must not be replaced 
by ordinary batteries.
CheckPoint should be cleaned with a mild 
detergent solution.

Operation

Start-up 

 (1)

Press the 

 key to switch on the CheckPoint. 

When the CheckPoint switches on, the version 
number of the installed software is shown briefly, 
followed by the result of the latest measurement 
(4).

Start/Stop measuring 

 (2)

Press 

 to start measuring. The built-in pump 

starts to suck measuring gas through the needle. 
The pump stops after 15 seconds, and the result 
is shown.

Extended measuring

If a measuring period of more than 15 seconds is 
required, it can be extended by pressing 

 

before the current measurement is complete. 
Each time you press 

, a 15 second measuring 

period is started.

Reduced measuring time

The measuring period can be reduced by 
pressing 

 while measuring is in progress.

Memory function 

CheckPoint remembers up to 10 measurements, 
including the latest one. The small display (6) 
shows the number of measured values stored. If 
the display is off, 9 measurements and the latest 
one are stored.

Clear memory

If the memory is full, it must be cleared before 
new values can be stored. Press 

 and keep it 

down. After 2 seconds the memory is cleared, 
and "0" is shown in the display (6), indicating 
that the memory is empty.

Show collected measurements

Press 

, and the first collected measurement is 

shown. Press 

 again to show subsequently 

collected measurements. While measurement 
values are being shown, the display (6) flashes, 
indicating the number of the collected 
measurement. To leave the display of collected 
measurements, press 

 until all measurements 

have been shown, or press 

 to return directly 

to the latest measurement. The display (6) stops 
flashing when you leave the "show collected 
measurements" function.

Replace collected measurement with new 
measurement

When showing the collected measurements (the 
display (6) is flashing), the measurement shown 
can be overwritten by a new measurement. While 
the collected measurement is shown, press 

 to 

start measuring. After measuring, the 
measurement is stored in the memory location 
shown.

EN

Summary of Contents for CheckPoint O2

Page 1: ...PBI 320297 B 02 2014 CheckPoint O2 DE DA SV RU FR ES IT EN PT...

Page 2: ...S COPYRIGHT PBI 320297 B 02 2014 CheckPoint O 2 Distributor Sales Service and Calibration Topac Inc 231 CJC Highway Cohasset MA 02025 Tel 781 7408778 Fax781 740 8779 Email sales topac com internet ww...

Page 3: ...2 CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S 7 13 5 14 6 4 1 3 2 7 8 8 9 11 12 10...

Page 4: ...condensing O2 Sensor Electrochemical Range 0 100 Resolution 0 1 Accuracy 0 25 abs 2 rel Temperature compensation 2 rel steady state Response time T95 15 sec Expected sensor lifetime 9 months Recommen...

Page 5: ...10 Brugerguide DA CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S NOTER...

Page 6: ...second measuring period is started Reduced measuring time The measuring period can be reduced by pressing while measuring is in progress Memory function CheckPoint remembers up to 10 measurements incl...

Page 7: ...n performing consecutive measurements in areas with large differences in the O2 concentrations the device will reach 95 of the true value during the first measurement 15 sec Example See illustration b...

Page 8: ...During calibration the display alternately shows CAL and 20 9 for 45 seconds CheckPoint has now been calibrated Stop calibration Press to interrupt the calibration process External power supply charg...

Page 9: ...t the needle retainer 9 in the opposite side and press it down until flush with the side plate 6 Test control Switch on the CheckPoint and start measuring with atmospheric air 20 9 Block the gas inlet...

Page 10: ...205 Needle kit 0 5 mm 100 pcs Ser Cpl 280167 Needle kit blunt 10 pcs Ser Cpl 310335 Filter sample gas 0 2 water trap 10 pcs Ser Cpl Spare parts 200720 Pump CheckPoint Ser Cpl 200721 Kit hoses internal...

Page 11: ...16 User Guide EN CheckPoint O2 PBI 320297 B 02 2014 COPYRIGHT Dansensor A S NOTES...

Reviews: