DANOPLUS DP-101 User Manual Download Page 2

TECHNICAL PARAMETERS

MAINTENANCE

1. Turn on the meter and carefully plug the microphone head in

    1/2 inch hole of calibrator

2. Turn on the calibrator and set the calibrator to output 94dB@1KHz.

3. Use flat screwdriver to adjust the calibration potentiometer

    located at the back, and make the LCD display 94.0.

4. Turn off the calibrator and remove the microphone very carefully.

The meter has been calibrated properly in the factory, the recommended 

calibration interval is 

1 year.

• Data recording & exporting

1. Data recording through APP on the phone

 (the data will be stored on the 

app and can store up to 100,000 groups).

   a. Click “Setting” in the “Menu”, then set the mode of reading, recording 

intervals, and alarm value to your liking and save the settings.

   b. Click “Record on Phone” on IOS version and “Start” on Android version 

to start the recording. Real-time data and data graph will appear.

   

c. Click “Stop” to stop the recording

and then click “Save” in the “Menu" to

save the data. You will now be asked

to name and save the file.

 

  d. To clear the data recorded on the

app, click “Clear” in the “Menu”.

Note:

 

Please must save the data first after finishing the recording. The 

data recorded on the app will be cleared up when you close the app.

2. Data recording on Meter

 (the data will be stored in the meter and can 

store up to 20,000 groups).

   a. Click “Setting” in the “Menu”, then set the recording intervals and the 

total number of records you wish to set for your meter

Note:

 

The meter will automatically stops recording when the saved data 

reaches the total number of records you set.

   b. To start the data recording, click 

“Start”

 on the app or press and hold 

the “       ” key on the meter until the dBA symbol starts flashing.

   c. To stop the data recording, click 

“Stop”

 on the app or press and hold 

the “       ” key on the meter until the dBA symbol stops flashing.

   d. Click 

“Export”

 on the app to export the stored data in the meter to the 

app. Set the name and then save the file.

   e. To clear the data recorded in the meter, click 

“Clear”

 on the app.

Click to start

measuring

Measured value

Data graph

- 5 -

- 4 -

- 7-

- 6 -

FOR IOS VERSION ONLY

• To View the data

   1. Click the “data file” on the top right corner of the interface.

   2. Click the “File name” to check the data.

• To share the data

   1. While on the Data file page, swipe

the file to the left. Share and Delete

icon will be shown.

   2. Click the “Share” icon, and can share on the

available apps.

FOR ANDROID VERSION ONLY

• To view the data

   1. Click the “data file” on the top right corner of the interface.

   2. Click the “file name” to check the data.

   3. Can be viewed as

       Curve data or List data.

• To export the data

1. While on the “record file”, click the dropdown button and select “Export”.

2. Choose to export via PDF or Excel file.

3. Click “Confirm” to download the file. The downloaded file will be stored 

in the "dB_Tester" folder located in the /storage/emulated/0/

The meter is powered by 3 pcs of AAA 1.5V Batteries.

Replace the batteries when:

    • The low battery signal icon         will appear on LCD.

    • The meter cannot be powered on.

To install the batteries:

    1. Turn off the meter and remove the battery cover.

    2. Replace the old batteries with 3 new AAA 1.5V batteries.

    3. Put back the battery cover.

1. If the device will not be used for a long period of time, remove the 

battery to prevent damage to the device due to battery leakage.

2. To prevent damage or severe deterioration to the device, do not open 

the inside of the device or modify its circuitry.

3. Do not place the device in a humid, dusty, acidic, alkaline or corrosive 

environment to avoid damage to the device.

4. Use a clean, soft cloth to wipe the meter if it becomes dirty.

Measuring Range

Accuracy

Frequency Response

Frequency Weighting

Resolution

Working Temperature and Humidity

Storage Temperature and Humidity

Power Source

Weight

External Dimension

Data Storage

Auto power off

30 ~ 130dBA

±1.5dB

31.5Hz ~ 8KHz

A Weighting

0.1dB

0 ~ 40°C, 10 ~ 80%RH

-10 ~ 60°C, 0 ~ 90%RH

3*1.5V AAA Batteries

96.38g (Excluding Battery)

56.1 x 177 x 36.03 mm

20000 groups

10 minutes without operation

BATTERY REPLACEMENT

VIEWING & SHARING SAVED DATA 

Install three 1.5V AAA batteries

Curve Data

Click the dropdown button

to choose between Curve

and List data view.

List Data

Save or share the file through the

application on the Share Sheet.

1

2

3

Swipe to the left

1

2

IOS version

Android version

IOS version

Android version

IOS version

Android version

CALIBRATION METHOD

To calibrate the meter, an external sound level calibrator that can provide a 

94.0dB signal at 1kHz and a small flat screw-driver are required. Follow the steps 

below to calibrate the instrument.

Data file

Data file

Menu

Menu

IOS version

Android version

IOS version

Android version

downloaded files location

Reviews: