Maintenance
Sintered Filters Replacement
For the correct function of the instrument, replace regularly the sintered filters in the pressure inputs.
Unscrew the pressure input using the 13 mm wrench, replace the clogged filters with a new set, screw
the pressure inputs back and tighten.
Battery Replacement
Unscrew the screws on the battery cover.
Remove the cover and insert new batteries following the technical parameters.
If you insert batteries of a different type, do not forget to change the battery type in the PFM 1000
application.