52
6
Configuration
Item
Technical Specification
TE-W2
Detec
tor
Type Un-cooled
FPA
micro-bolometer
Array size
384*288
Pixel Pitch
25
μ
m
Spectral range
8-14
μ
m
Frame rate
50Hz
Thermal
sensitivity
0.06
℃
@30
℃
Lens
FOV 25°×19°
Min Focus
Distance
10cm
Focus Manual
Interface Type
Screw Type
Imag
e
IFOV 1.36mrad
Zoom
2×
、
4X
CCD
3 million pixel
,
CMOS module
Frame Rare
50HZ/60HZ
,
Video Output
PAL / NTSC
Imag
e
Displ
ay
LCD display
3.6
″
TFT LCD
,
640*480
,
touch screen
Color Palette
11 color palette for option
Image
adjustment
Auto/manual to adjust brightness, contrast
Image
Fast switch between infrared image and ccd image, with
fusion function.
Image Enhance
Auto/manual enhance mode
Meas
urem
ent
Range
+30
℃
~+42
℃
Accuracy
± 1 °C
Measurement
Correction
Auto / Manual
Mode
Up to 5 movable spots. Up to 5 movable areas
(maximum, minimum and average temperatures). Up
to 2 movable lines. Line profile. Isotherms. Temperature
difference. Alarm(voice, color)
Emissivity
correction
Variable from 0.01 to 1.0
Background Tem
correction
Auto
Atmospheric
transmission
correction
Automatic correction according to user input object
distance, humidity and temperature
Setting Option
Time/Date,
℃
/
℉
/K, language, brightness/contrast, LCD
brightness, IP address
Imag
Card
8GB Micro TF card
,
max 32GB