DX-1 Ultrasonic Micrometer
13
At this point, the
DX-1
has successfully calculated it's internal error
factor, and will compensate for this value in any subsequent
measurements. When performing a "probe-zero", the
DX-1
will always use
the sound-velocity value of the built-in probe-disc, even if some other
velocity value has been entered for making actual measurements. Though
the
DX-1
will remember the last "probe-zero" performed, it is generally a
good idea to perform a "probe-zero" whenever the tool is turned on, as well
as any time a different transducer is used. This will ensure that the
instrument is always correctly zeroed.
Calibration
In order for the
DX-1
to make accurate measurements, it must be set to
the correct sound-velocity for the material being measured. Different types
of material have different inherent sound-velocities. For example, the
velocity of sound through steel is about 0.233 inches-per-microsecond,
versus that of aluminum, which is about 0.248 inches-per-microsecond. If
the tool is not set to the correct sound-velocity, all of the measurements the
tool makes will be erroneous by some fixed percentage. The
one point
calibration is the simplest and most commonly used calibration procedure -
optimizing linearity over large ranges. The
two point
calibration allows for
greater accuracy over small ranges by calculating the probe zero and
velocity. The
DX-1
provides three simple methods for setting the sound-
velocity, described in the following pages.
Summary of Contents for DX-1
Page 4: ...Dakota Instruments blank page ...
Page 6: ...Dakota Ultrasonics 2 blank page ...
Page 18: ...Dakota Ultrasonics 14 ...
Page 27: ...DX 1 Ultrasonic Micrometer 23 blank page ...
Page 31: ...DX 1 Ultrasonic Micrometer 27 blank page ...