DAKOTA ULTRASONICS Z-302-0001 Operation Manual Download Page 26

 

23 

 

CHAPTER SIX 
ADDITIONAL FEATURES 

6.1 Gain 

The gain, or amplification of the return echoes, can be adjusted in the 

ZX-3

 to 

accommodate a variety of materials and applications.  The setting of the gain is 
crucial in order to obtain valid readings during the measurement process.  Too much 
gain may result in erroneous measurements, detecting on noise rather than the 
actual material back wall surface.  Not enough gain may result in intermittent 
detection.  It could also result in lack of detection on internal flaws, pits, or porosity. 
The gain can be compared to the volume control of a home stereo system.  If you 
turn it up too much, you can’t hear the music clearly.  If it’s turned down too much, 
you can’t hear it at all. 
 
The

 ZX-3 

has five gain settings 

(VLOW, LOW, MED, HIGH, VHI)

.  The gain range is 

40dB – 52dB in 3dB increments.  The 

ZX-3

 has been optimized for the 

MED

 gain 

setting at 46dB for all common applications.  It should be operated in this mode as 
standard.  However, some applications may require the lower or higher gain settings.  
When?  The low settings may be necessary for noisy or granular cast materials.  How 
do I know when to lower the gain?  If the reading becomes sporadic and won’t settle 
down or resolve on a thickness value because the material is either very noisy or 
granular.  Setting the gain to a lower less sensitive level, would potentially offer 
improved stability. 
   
How do I know when to increase the gain?  When a material is difficult to penetrate or 
pass sound through.  This could be due to the material type, overall thickness, the 
transducer diameter and frequency, or a combination of all the above.  Turning the 
gain up for additional output could improve the ability to obtain a successful 
measurement.  Another example would be the need to increase overall sensitivity for 
locating fine pits or flaws.  In any case, the selectable gain settings offer improved 
versatility to resolve and overcome potential application issues. 

 

The procedure for editing the gain is outlined as follows: 
 

GAIN 

 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for Z-302-0001

Page 1: ...OPERATION MANUAL DAKOTA ULTRASONICS Z ZX X 3 3 Ultrasonic Thickness Gauge P N P 302 0002 Rev 1 10 March 2019 www com information itm com 1 800 561 8187...

Page 2: ...LTRASONIC MEASUREMENT 9 3 1 TIME VERSUS THICKNESS RELATIONSHIP 9 3 2 SUITABILITY OF MATERIALS 9 3 3 RANGE OF MEASUREMENT AND ACCURACY 9 3 4 COUPLANT 9 3 5 TEMPERATURE 10 3 6 MEASUREMENT MODES 10 CHAPT...

Page 3: ...6 6 PROBE DIAMETER 27 6 7 LOCK 28 6 8 FACTORY DEFAULTS 29 APPENDIX A VELOCITY TABLE 31 APPENDIX B APPLICATION NOTES 33 www com information itm com 1 800 561 8187...

Page 4: ...material being measured 1 1 Disclaimer Inherent in ultrasonic thickness measurement is the possibility that the instrument will use the second rather than the first echo from the back surface of the m...

Page 5: ...o a computer keyboard This key will be used to select or set a menu option Note Unit will automatically power off when idle for 5 minutes All current settings are automatically saved prior to powering...

Page 6: ...erials and applications There are five gain settings VLOW LOW MED HIGH VHI The gain range is 40dB 52dB in 3dB increments Refer to page 23 for details regarding the gain feature 2 5 SCAN Key The ZX 3 o...

Page 7: ...ting through the various features in MENU Navigating the Features in Menu 1 Press the key once to enter the sub menu items 2 Press the keys to toggle through the features 3 To enable or edit the statu...

Page 8: ...to change a value or setting B Large 7 segment The thickness measurement velocity or alpha message will be displayed in this area C Measurement Modes This icon indicates the measurement mode The ZX 3...

Page 9: ...custom material type L Features The icons illuminated in this row across the bottom of the LCD display which features are currently enabled For a complete list of the menu features in the ZX 3 Refer...

Page 10: ...accomplished with the use of a coupling fluid commonly called couplant This fluid serves to couple or transfer the ultrasonic sound waves from the transducer into the material and back again Before a...

Page 11: ...disk are located on the ZX 3 s top end cap The transducer connectors are of type Lemo 00 Note There is no polarity associated with connecting the transducer to the ZX 3 it can be plugged into the gaug...

Page 12: ...tical in a wide variety of materials including metals plastics and glass Materials that are difficult include some cast materials concrete wood fiberglass and some rubber 3 3 Range of measurement and...

Page 13: ...s 3 6 Measurement Modes This section will cover the different measurements modes of the ZX 3 the transducers required and the reasons for using specific modes Pulse Echo P E Mode Pulse echo mode measu...

Page 14: ...material frequency and diameter A limitation of dual element delay line transducers is the V shaped sound path Because the sound travels from one element to another the time versus thickness relations...

Page 15: ...still requiring the ability to detect flaws and pits A special coating style transducer is required for use in this mode There are a variety of coating transducers in various frequencies available fro...

Page 16: ...ce noise issues or mirroring effect Higher frequency transducers 7 10MHz offer improved resolution to avoid erroneous measurements Restricted access Measuring materials with extreme curvatures or rest...

Page 17: ...o determine the velocity of a single sample refer to the Material Calibration section on page 16 When measuring curved materials it s more accurate to calibrate from two test points one at the minimum...

Page 18: ...d flashing CLn Prb clean probe 2 Make sure the couplant is wiped clean from the tip of the transducer to avoid any zero error 3 Press the key to perform the zero Performing a Manual Probe Zero On Bloc...

Page 19: ...econd If the gauge is not set to the correct sound velocity all of the measurements the gauge makes will be erroneous by some amount The One Point calibration is the simplest and most commonly used ca...

Page 20: ...point calibration option The one point calibration option is most suited for linearity over large ranges When using the one point option the calibration should be perform on the thickest side of the m...

Page 21: ...wn thickness value Note The longer the keys are pressed and held the faster the value will increment decrement Note Pressing the key prior to pressing the key will abort the cal routine without saving...

Page 22: ...he probe zero standard For clarification if it s desired to use a one inch reference of a specific material type as the zero reference performing the first point of a two point calibration sets the in...

Page 23: ...easurement screen Note The internal zero of the ZX 3 is now set The procedure above can be used to set the internal zero of the ZX 3 to use any reference standard as the probe zero standard if desired...

Page 24: ...offer a reasonable approximation of the thickness There are 9 common materials and 2 user programmable settings available It s important to note that these velocities will not always be an exact repre...

Page 25: ...ill abort to the measurement screen without saving any changes 4 If USR1 or USR2 were selected the velocity edit screen will be displayed and edit icon illuminated and flashing 5 Use the keys to scrol...

Page 26: ...operated in this mode as standard However some applications may require the lower or higher gain settings When The low settings may be necessary for noisy or granular cast materials How do I know when...

Page 27: ...can feature of the ZX 3 increases the overall repetition rate to a maximum of 140Hz with a high speed screen refresh rate of 25 times a second This allows for making scanned passes over an arbitrary l...

Page 28: ...ressing the key prior to pressing the key will abort to the measurement screen without saving changes 5 When the desired UNIT setting is displayed press the key to set the units and return to the meas...

Page 29: ...gh the BRT LO MED HI options 7 When the desired BRT setting is displayed press the key to set the brightness and return to the measurement screen Note Pressing the key prior to pressing the key will a...

Page 30: ...ting is displayed press the key to set the status and return to the measurement screen 6 6 Probe Diameter The PROB feature was added to improve linearity when using a specific probe diameter The defau...

Page 31: ...s displayed press the key to set the diameter and return to the measurement screen 6 7 Lock The lock feature was built into the ZX 3 for the purpose of locking the operators out of editing any of the...

Page 32: ...ot functioning properly or perhaps multiple features have been enabled and a clean start is needed The procedure to reset the gauge is outlined below Factory Reset 1 With the ZX 3 powered off press an...

Page 33: ...to edit the reset option 6 Press the keys to toggle YES followed by pressing to reset the gauge 7 Repeat the steps above to set MEDI ZERO back to their original settings noted in step three above www...

Page 34: ...2290 5817 Iron 0 2320 5893 Cast Iron 0 1800 approx 4572 Lead 0 0850 2159 Magnesium 0 2300 5842 Mercury 0 0570 1448 Molybdenum 0 2460 6248 Monel 0 2110 5359 Nickel 0 2220 5639 Nylon 0 1060 approx 2692...

Page 35: ...Dakota Ultrasonics 32 Tin 0 1310 3327 Titanium 0 2400 6096 Tungsten 0 2040 5182 Uranium 0 1330 3378 Water 0 0580 1473 Zinc 0 1660 4216 Zirconium 0 1830 4648 www com information itm com 1 800 561 8187...

Page 36: ...eratures above this point the change in sound velocity of the material being measured starts to have a noticeable effect upon ultrasonic measurement At such elevated temperatures it is recommended tha...

Page 37: ...option An additional important consideration when measuring laminates is that many included air gaps or pockets which will cause an early reflection of the ultrasound beam This effect will be noticed...

Page 38: ...t to this product Dakota Ultrasonics specifically disclaims any implied warranties of merchantability or fitness for a particular purpose Some states or provinces do not allow limitations on the durat...

Reviews: