paused.  And, you need  to  release  the  pause  of each  operation  mode  by  pushing  the  jog 

shuttle if you want to restart overall timer operations.

5. Locking Mode

After completion of set up the temperature and timer, you can set the Locking (Safe) Mode by 

clicking  the  mode  button.  In  the  Locking  Mode,  the  LCD  shows  the  current  temperature  in 

measuring value digit. In this mode, all inputs are disabled using the jog shuttle. If you want 
to release this mode, click the mode select button. 

6. Temperature Offset 

Offset Function  is  addition  or  subtraction  of indicated  temperature.  It  is  very  convenient 

function  to  synchronize  displayed temperature  of  the  plate  and  sample with  the  measured 
value by the User.

A. Press Mode Switch and Jog Shuttle simultaneously in the Heating Mode (Mode S/W + Jog 

S/W).

B.  In  this  status,  it  is  changed  into  “Offset”  Mode  and  the  user  can  adjust  the  value by 

pushing the knob once, then the offset would be setup.

C. After the setup, the temperature on LCD display indicates the compensated temperature.
D.  At  this  time,  press  the  mode  switch  and  jog  shuttle  simultaneously,  then  escape  from 

Offset Mode.

7. Storage of the set value

Digital  Feedback  controller  have a  “Memory”  function  to  save  each set  value.  The user  can 
save  the  set  values  to  be  used  frequently  and  does  not  have  to  set  the  same  value  every 
moment.

A. Set the value of each mode (except Locking mode) using Jog Shuttle.

Press the knob over one second, then the set value is blinking on LCD three times and it 
means that the value has been saved.

8. Power off

After you complete to use the device, you must turn off the power switch for your safety (

).

Summary of Contents for WHM-C10D

Page 1: ...ISO 9001 Certified PL Insurance Patented No 10 0554098 Jog Shuttle Control System Operation Manual Version 2 1 1 written for Model WHM C10D WHM C10D Models DAIHAN Sci com or ALL for LAB com Shopping M...

Page 2: ...ler WHM C10D to which this declaration relates is in conformity with the following standard or other normative document EN 61326 EN 61000 3 2 EN 61000 3 3 EN 61010 1 EN 61010 2 following the provision...

Page 3: ...IME Jog Shuttle switch Turn Push l User s self compensation function to control difference between Real Temp and l Indicated Temp 10 0 l Digital LCD with Back Light Function l Storage function Timer A...

Page 4: ...Mall DAIHAN Scientific WHM C10D Digital Feedback Temp Controller 4 27 Appearance 3 The appearance of Digital Feedback Controller WMH C10D is shown in Figure 1 Figure 1 Appearance of WMH C10D Figure 2...

Page 5: ...Multifunctional Digital Feedback Controller Figure 2 Controller of Digital Feedback Controller Operation Status on LCD Display Indication of Set Value Indication of Current Value LED indicating Heatin...

Page 6: ...sage of the external sensor SS300 optional K type sensor or SS310 optional Teflon coated K type sensor A If you don t connect K type sensor the temperature is controlled by the measured value of Heati...

Page 7: ...shuttle for adjustment of temperature the set value is change firstly and the current value goes after the set value Then the current value is reached to the set value and it means that Digital Feedba...

Page 8: ...nd sample with the measured value by the User A Press Mode Switch and Jog Shuttle simultaneously in the Heating Mode Mode S W Jog S W B In this status it is changed into Offset Mode and the user can a...

Page 9: ...ating and touching this surface may cause a serious injury If you need to touch it you should have sufficient cooling time down to 70 on the LCD display before touching The recommended power supply is...

Page 10: ...ct is designated for separate collection at an appropriate collection point Do not dispose of as household waste This product is required to comply with the European Union s Waste Electrical Electroni...

Page 11: ...unctions before shipment 2 This instrument is warranted for two years from purchasing date 3 A breakdown caused by customer s misuse or natural disaster is excepted from this warranty 4 This certifica...

Reviews: