DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
DAIHAN Scientific, Forced Convection type Smart Incubator 34/81
<Fig.3.3.27> check temperature graph
<Fig.3.3.28> change display scope
-
Wait time can be set to 23 hours and 59 minutes and Temperature Zone can be set to
±
10
℃
.
-
If the target temperature does not reach Temperature Zone within Wait Time, it moves
on to the next Segment.
3.3.3. CHART
-
Temperature data is stored at internal memories at every 10 seconds and maximum 6
months of data are stored. Please use Data Backup once every 6 months.
-
Temperature scope shown on graph can be changed using the set button on the low left.
Maximum and minimum values can be changed within limits shown on a pop up window.
-
-
After inserting a USB into the port on top of controller, it stores in the USB if the store icon
on top right is pressed.
-
Error message appears if USB is not recognized. Press “OK” button and remove USB and
reinsert and press the store icon.
Upper temperature range
setting button
Lower temperature range setting button
Save