3. Product Usage

      3.3 Usage Instructions

-

Before operation, Set the SAFETY knob to 30~40

°

C higher than operation temperature.

-

Connect the power cord to the proper power socket and Turn on the Main Switch.

* Operation method for Jog-Dial

-

Jog-Dial can be manipulated for the setting in each mode by Turning and Push motion.

-

If  turning  the  Jog-Dial  knob  to  clockwise  or  counter-clockwise,  the  setting values  can  be 
adjusted, and if briefly pressing once the PUSH  button, the Start or the Stop operation is 
performed.

-

The  Turning  of  the  Jog-Dial  knob  does  not  manipulated  during  operating,  and  if resetting 
the set value, it can be adjusted by the Jog-Dial knob after stopping the product by pressing 
briefly once the PUSH button.

-

If  pressing  for  2  or  3  seconds  the  PUSH  button,  the  set  value  is  saved,  and  even  if  the 
power is turned on again, that value is maintained. (If “0” value is saved, the setting value 
is initialized.)

MAIN - Set Operation Time 

-

When the power is turned ON, the initial screen is the setting mode of the Timer set mode.

                           

-

TIMER LED is lighted during the operation of timer, and indicates that it is operating.

-

The range of setting time is from 00:00 to 99hr and 59min(it is possible to adjust at 1min.)
and 00:00 indicates continuous operation.

-

If the timer operation is completed, the alarm will sound and heating will be stopped.

-

The timer setting mode can be operated by referring to the jog-shuttle operation method.

< Fig 3.3.1. Set Operation Time Mode>

Time  SV    :            00:00
Time  PV    :            00:00

Set Time

Remain Time

Timer LED (Green)

Summary of Contents for ThermoStable OF-50

Page 1: ...Liability Insurance Patented No 10 0554098 Jog Shuttle Control System Validation IQ OQ Service User Manual Version2 2 1 written for ThermoStableTM OF 50 105 155 W105 W155 models ThermoStableTM OF 105...

Page 2: ...DAIHAN Sci com or ALL forLAB com Shopping Mall DAIHAN Scientific Forced Convection type Oven 2 48...

Page 3: ...Product Introduction 5 1 3 Product Configuration 6 2 Installation 2 1 Product Installation 7 2 2 Product Specifications 9 3 Product Usage 3 1 Product Usage 10 3 2 Names of Each Part 13 3 3 Usage Inst...

Page 4: ...ing Warning Failure to follow the instructions may result in injury Caution Failure to follow the instructions may result in product failure Proximity Caution Keep clear of the product during use Noti...

Page 5: ...utting the Air Flow Layer between the advanced insulator and the enclosure 6 The Chamber and inner shelf prevents corrosion by using Stainless Steel material 7 This product minimizes the heat loss by...

Page 6: ...Liter ThermoStable OF 155 General Forced air Oven without Window 155Liter ThermoStable OF W105 General Forced air Oven Built in Viewing Window 105Liter ThermoStable OF W155 General Forced air Oven Bui...

Page 7: ...mal operation at a rated voltage so please check the voltage status before installing the product 7 When installing and using the product please certainly use the power cord provided with the products...

Page 8: ...ols stickers Attached to side of product For meaning of each symbol please refer to Precautions Before Use Safety Instructions Product Liability PL Insurance Admission DAIHAN Scientific Co Ltd is a me...

Page 9: ...min to 150 6min to 100 6min to 150 Controller Digital Fuzzy Control by Advanced Microprocessor Jog Dial with Push Button RS232 Port Available to connect with PC Display Digital LCD with Back Light Tim...

Page 10: ...distress the power cord Please do not hurt on the power cord or process it or force bend it or pull it and tie it Also if heavy objects are placed over the power cord or the power cord is inserted in...

Page 11: ...hines and can get rained on it is prohibited It may cause overheating or electrical fault of the product Please do not use flammable or volatile products in the vicinity of the product to minimize the...

Page 12: ...d please contact the deliverer or our technical service engineer Do not operate the unit before checking with an engineer otherwise it may cause an electrical shock or a fire The Sound Level of the bu...

Page 13: ...tableTM OF 105 155 W105 W155 2ea Controller Set and Check operation Window Safety Double Door Handle Door open and close Fig 3 2 3 Back of ThermoStableTM OF 105 155 ThermoStableTM OF W105 W155 Fig 3 2...

Page 14: ...OFF Fuse Protect unit from High Current Safety Switch Prevent overheat RS232 connection Serial Interface for Remote control by PC Fig 3 2 5 Standard Model ThermoStableTM OF 105 Fig 3 2 6 Side of Ther...

Page 15: ...ation of TIMER HEAT ERROR REF MAIN Button Convert into mode of Timer Heating Locking SUB Button Convert into mode of delay timer and offset function Jog shuttle knob Control that Reduces or Adds Value...

Page 16: ...it can be adjusted by the Jog Dial knob after stopping the product by pressing briefly once the PUSH button If pressing for 2 or 3 seconds the PUSH button the set value is saved and even if the power...

Page 17: ...d the reservation timer is activated After setting the reservation timer value you can set up the operation timer value by pressing the MAIN button and then press Start button And you can set up the t...

Page 18: ...ndicates the standby status while blinking in long interval The range of the setting temperature is adjustable in the unit of 1 and can be set up to 230 You can change or save the setting by using the...

Page 19: ...king If pressing MAIN button in the Set Temperature Mode it is switched to the Locking mode In the Locking Mode the Turning by the Jog Dial Knob or Push button do not applied and LCD Display and LED l...

Page 20: ...1050 Validation Service lQ OQ OFV1050 for ThermoStable OF 50 DH WOF31105 Validation Service lQ OQ OFV1105 for ThermoStable OF 105 DH WOF31155 Validation Service lQ OQ OFV1155 for ThermoStable OF 155 D...

Page 21: ...t and soft cloth is recommended Do not use coarse cloth strong chemicals or organic solutions 3 Do not use water Benzene Thinner or any alcohol for cleaning the product It may cause discoloration dama...

Page 22: ...f current temperature of product when slowly turning clockwise up to 320 after adjusting the Safety Knob to 0 4 Tak sound is Normal If Tak sound is not occurred A S is required 5 Please check the prog...

Page 23: ...ted Heating but temperature is fluctuated Check the temperature that is currently set Make sure that the power switch is in the ON position Check that SAFETY Knob is not placed in the 0 position and i...

Page 24: ...ument is warranted for 2 years from purchasing date 3 A breakdown caused by customer s improper use or natural disaster is excepted from this warranty 4 This certification is only for international cu...

Reviews: