DAIHAN-Sci.com or
ALL
for
LAB.com
Shopping Mall
High Performance Dry Bath Heating & Cooling
20/48
4. Temperature Setting Mode.
●
SV, the value of first line (smaller-sized digits) on LCD display, indicates set value of operation
Temperature and PV, the value of second line (bigger-sized digits) indicates present value
of the Temperature.
●
Turn the Jug shuttle Knob such as clockwise or counterclockwise then SV and PV change
together. SV can be set in 0.1
℃
.
●
Temperature Range : HB-48 / HB-96 / HB-96D
→
A5
℃
~ 150
℃
HB-R48
→
-5
℃
~ 95
℃
●
When you finish Set Value, push Jog shuttle Knob 1 time shortly, the unit will be running
and Heating LED(Red) will be turn on after 2~4 seconds.
●
When you finish Set Value, push Jog shuttle Knob About in 2~3 seconds, will be save the
Set Value of Temperature.
●
After that, push the Jog shuttle Knob 1time shortly, the heater will be running. After 2~4
seconds, you will see “Heating” message on the LCD display.
●
it cannot re-set the Set Value of Temperature when the heater is running, if you want to re-
set the Set Value of Temperature, push the Jog shuttle Knob 1 time shortly (Pause mode) and
set to re-set the Set Value of Temperature.
●
after re-set the set value of Temperature, push the Jog shuttle Knob in 2~3 seconds or
push the Jog Shuttle Knob 1 time shortly.
●
If you want to go to next mode(Safety Mode), push MODE button 1 time.
5. Safety Mode(Locking Mode).
●
In this mode, turning or pushing Jog shuttle knob doesn’t work for any set value.
So, you could prevent any changes of set values by mistake.