1. What is blackbody? What is the role? Do you need a physical connection to the camera? 

Blackbody can be simply understood as a constant temperature reference source (constant temperature can 

be maintained after power on). When testing human body temperature, blackbody is generally set at 35 

which  is  used  for  temperature  correction  of  thermal  imaging  acquisition,  so  as  to  meet  the  accuracy 

requirements of 

±

  0.3 

Bold  doesn't  need  to  be  connected  to  the  camera,  it's  separate.  After  power  on,  it  can  be  placed  in  the 

camera screen according to the installation instructions. 

2. Power on stabilization time of equipment. 

Blackbody: power on and preheat for 20 minutes; thermal imaging equipment: power on and preheat for 60 

minutes, with the best effect. 

3. Whether it can be installed in outdoor or semi outdoor environment. 

Because the outdoor temperature, wind, humidity and other environmental factors have a great impact on 

the  temperature  measurement  of  the  thermal  imaging  surface,  it  is  not  recommended  to  be  installed 

outdoors or in the area directly connected to the outdoor, but in a closed and relatively stable environment 

without wind. 

4. Why is it recommended to measure the temperature of frontal forehead? 

First of all, different people's forehead temperature also has temperature difference. In the case of mask, the 

exposed area of forehead is relatively large, which is more convenient for temperature measurement. 

The actual temperature of the front face and the side face of the same person / same position is different. 

With masks, the high temperature points are mainly in the forehead, ears and neck, and the temperature is 

different  among  different  people.  Therefore,  it  is  required  to  face  the  camera  and  test  the  frontal 

temperature uniformly to reduce the impact of temperature difference between different parts. 

5. What is the best temperature range for temperature measurement? 

The best temperature range of indoor wind free and stable environment is 10 

  ~ 30 

 

6. The best distance for temperature measurement? 

The best temperature measurement distance is the same as the distance between blackbody and equipment. 

3.5mm detection distance is 2m, and the transverse width of 2m is about 1.5m; 7mm detection distance is 

3M, and the transverse width of 3M is about 1.3m; the temperature measurement value far from the optimal 

distance  will  be  reduced,  and  the  temperature  measurement  value  close  to  the  optimal  distance  will  be 

increased. 

7. How many heads can be tested? 

As  shown  in  the  figure  below,  in  normal  use,  there  are  not  many  people  (width  1.3-1.5m)  at  the  best 

distance of temperature measurement. There is no limit to the number of heads in the detection area, but it 

is  too  far  or  there  is  overlap,  which  may  result  in  missing  or  inaccurate  temperature  measurement.  It  is 

recommended to test one by one. 

 

8. Why does the image of thermal imaging channel get stuck after a period of time? Is it normal? 

Normal.  In  order  to  ensure  image  uniformity  and  temperature  measurement  accuracy,  thermal  imaging 

 

 

Summary of Contents for BF Series

Page 1: ...Installation and commissioning guide for temperature measurement for human body of Dahua thermal camera BF Series Zhejiang Dahua Technology Co Ltd...

Page 2: ...ture exceeds 35 there is a possibility of fever For other environmental temperature conditions it is recommended to measure several samples on site and increase 1 as the alarm threshold on the basis o...

Page 3: ...Preface II Change record Revision date Version number Creater Reviser 2020 01 27 V1 0 2020 01 28 V1 1 2020 01 31 V1 2 2020 02 03 V1 3...

Page 4: ...tallation precautions 2 General settings 7 2 1 Device upgrade 2 2 Image debugging 2 3 Human body temperature detection function debugging 2 3 1 Select the intelligent scheme of human body temperature...

Page 5: ...n of installation guidance video web operation guidance and guidance document should be obtained from the regional delivery supervisor In the future it will be synchronized in the group on the mobile...

Page 6: ...stallation requirements are consistent with bf3221 7mm 1 3 Installation diagram Bf3221 3 5mm installation diagram Blac kbod y Height 2m Side view h u m a n Distance between two objects 2m Height of Bl...

Page 7: ...istance between two objects 3m Height of Blackbody 1 8m Thermal camera focal length 13mm Blac kbod y Sidewalk width 1 5m Top view Thermal camera focal length 13mm 1 4 Real scene diagram BF3221 7mm as...

Page 8: ...ween the camera and blackbody Thermal camera Height 2m Best test distance 3M Height of Blackbody 1 8m Personnel access direction 1 When the artificial diversion channel is lengthened as far as possibl...

Page 9: ...r right side of the thermal imaging picture and the tripod can be protected by temporary fence As follows black body is in the upper right corner of the image It is required that the visible light cha...

Page 10: ...upgraded item No 2 6 01 05 18388 If there is a new iterative process will be notified in the delivery group please pay attention to the group information in time The default to be restored after upgr...

Page 11: ...ent scheme 2 2 2 Blackbody parameter setting Blackbody calibration Note the blackbody parameter setting steps are ignored for the scheme without blackbody a The actual setting of blackbody is 35 b Aft...

Page 12: ...est temperature is within 35 0 2 on average 2 2 3 Drawing of human body temperature detection area a BF3221 7mm BF5401 13mm i The subject stands on the edge of the blackbody 3m away from the camera fa...

Page 13: ...best temperature measurement distance and the measured temperature is generally high or low they can be compensated by temperature correction For example if it is 0 5 lower fill in 0 5 for temperatur...

Page 14: ...11...

Page 15: ...uracy or scrap so as to avoid the scratch of blackbody radiation surface and affect the temperature measurement accuracy Only for indoor use The use environment shall be free from obvious air convecti...

Page 16: ...mperature measurement The actual temperature of the front face and the side face of the same person same position is different With masks the high temperature points are mainly in the forehead ears an...

Page 17: ...ffect is not ideal please provide the Dav video with two channels on site Send Xu Fangrong and Fang Shengdong through Dahua interactive platform select intranet Download for outgoing type After receiv...