Information on EMC
7
Phenomenon
Basic EMC
standard or test
method
Operating
mode
Port tested
Test Voltage
Test level/requirement
Mains terminal
disturbance
voltage
CISPR11:2015
Operating
AC Mains
AC 120V,60Hz
AC 220V,60Hz
AC 230V,50Hz
Group1, Class A
Radiated
disturbance
CISPR11:2015
Operating
Enclosure
AC 120V,60Hz
AC 220V,60Hz
AC 230V,50Hz
Group1, Class A
Harmonic Current
Emission
EN 61000-3-2:2014
IEC 61000-3-
2:2014
Operating
AC Mains AC 230 V~,50 Hz
Class A
Voltage change,
Voltage
fluctuations and
Flicker Emission
EN 61000-3-3:2013
IEC 61000-3-
3:2013
Operating
AC Mains AC 230 V~,50 Hz
Pst: 1
Plt: 0.65
Tmax:0.5
dmax: 4%
dc: 3.3%
Electrostatic
Discharge
Immunity
EN 61000-4-2:2009
IEC 61000-4-
2:2008
Operating
Stand by
Enclosure
AC 120V,60Hz
AC 220V,60Hz
AC 230V,50Hz
± 8 kV/Contact
± 2, ± 4, ± 8,
± 15 kV/Air
Radiated RF
Electromagnetic
Field Immunity
EN 61000-4-
3:2006+A2:2010
IEC 61000-4-
3:2006+A2:2010
Operating
Stand by
Enclosure
AC 120V,60Hz
AC 220V,60Hz
AC 230V,50Hz
3 V/m
80 MHz-2.7 GHz
80% AM at 1 kHz
Immunity to
Proximity Fields
from RF wireless
Communications
Equipment
EN 61000-4-
3:2006+A2:2010
IEC 61000-4-
3:2006+A2:2010
Operating
Stand by
Enclosure
AC 120V,60Hz
AC 220V,60Hz
AC 230V,50Hz
Table 9 in
IEC 60601-1-2:
2014
Electrical Fast
Transient/Burst
Immunity
EN 61000-4-4:2012
IEC 61000-4-
4:2012
Operating
Stand by
AC Mains
AC 120V,60Hz
AC 220V,60Hz
AC 230V,50Hz
± 2 kV, 100 kHz
repetition frequency
Surge Immunity
EN 61000-4-5:2014
IEC 61000-4-
5:2014
Operating
Stand by
AC Mains
AC 120V,60Hz
AC 220V,60Hz
AC 230V,50Hz
Line to Line
± 0.5 kV, ± 1 kV
Line to Ground
± 0.5 kV, ± 1 kV, ± 2
kV
27