8
SETUP OF TECHNICAL
PARAMETERS
SET mode can be accessed, as shown in Fig. A, by
pressing the <Mode> button from DIVE or SURF
mode, or by pressing the <-> button from CLOCK
mode.
SET mode is subdivided into five function groups,
SET1.... SET5, as shown in Fig. B. It is possible to
navigate through the different SET function groups
in the following manner:
To enter a given SET function group, press the
<Mode> button for approximately 2 sec.
Alternatively, press and immediately release the
<Mode> button to skip to the next SET function.
Press and immediately release the <-> button to
go back to the preceding SET function.
From SET5, pressing the <Mode> button saves any
changes made to the settings and goes on to
CLOCK mode. The parameters within a given SET
function group are modified using the <-> button.
SET1 - S
ETUP OF
"A
IR
", "N
ITROX
",
"B
OTTOM
T
IME
"
OPERATING MODE
AND ASSOCIATED PARAMETERS
The display initially shows the following
parameters, (fig. 2):
"Nitrox" icon.
"Air".
"Gauge" (botton time).
FIG. 2
To enter SET1 press the < Mode > button for
approximately 2 sec.
Alternatively, to advance to SET2, press and
immediately release the <Mode> button .
O
PERATING MODE SETUP
: "A
IR
", "N
ITROX
",
"B
OTTOM
T
IME
"
The display shows the currently active operating
mode. Press the <-> button to change the
setting.
If 'Nitrox' mode is selected, pressing the
<Mode> button will advance to the setup of %
O
2
and then of max PPO
2
. Otherwise, if "Air" or
"Bt" mode is selected, pressing the <Mode>
button advances to SET2.
"N
ITROX
"
BREATHING MIX SETUP
The percentage of O
2
in the breathing mix can
be entered in the range between 21% - 50%, in
increments of 1% (fig. 3). Press the <-> button
to increase the percentage. Press the <Mode>
button to save the percentage value and
advance to the next parameter.
FIG. 3
!
WARNING
It is essential to correctly set the
oxygen percentage in the breathing mix
to ensure correct readouts of:
- no decompression time remaining
- decompression stop times
- alarm for exceeding maximum
permitted PPO
2
.
!
WARNING
Do not use mixes with an O
2
% greater
than 50%.
!
WARNING
If you have completed a NITROX dive and
are planning a repetitive dive using
compressed air, set your dive computer
for NITROX mode with a % O
2
of 21%. In
this way, the monitoring of the % CNS O
2
will remain active.
Darwin.qxd 29/03/2002 16.45 Pagina 8
Summary of Contents for DARWIN
Page 24: ...136 Darwin qxd 29 03 2002 16 47 Pagina 136...
Page 25: ...137 FIGURES FIG A Darwin qxd 29 03 2002 16 47 Pagina 137...
Page 26: ...138 Darwin qxd 29 03 2002 16 47 Pagina 138...
Page 27: ...139 FIGURES FIG B Darwin qxd 29 03 2002 16 48 Pagina 139...
Page 28: ...140 FIG C Darwin qxd 29 03 2002 16 48 Pagina 140...
Page 29: ...141 FIGURES FIG D Darwin qxd 29 03 2002 16 48 Pagina 141...
Page 30: ...142 FIG E Darwin qxd 29 03 2002 16 48 Pagina 142...
Page 31: ...143 FIGURES FIG F Darwin qxd 29 03 2002 16 48 Pagina 143...