dacell AM-1000 User Manual Download Page 8

 

7

 

3-3. Description 

 

1

  Symbol (-) display   

This symbol lamp will be lighted when the output voltage is become less than 

0V. 

  ○

2

  Voltage display 

Output voltage is displayed as in 4 Digit -7 Segment LED.   

When  output  voltage  is  over  10V  or  below  -10V,  it  will  be  displayed  as  +/-

10.00 and the rest section will be displayed as in +/- 9.000V.   

  ○

3

  Zero Hand-operated control volume 

It is a volume to control the offset of residual quantity

 

after finishing the auto 

zero control.   

Control range is within 

±

 1V. 

4

  Zero auto control switch   

Use this switch to carry out auto zero control.   

When  this  auto  switch  is  pressed,  “Auto”  will  appear  on  the  LED  and  then 

zero will be set. 

 

6

  Bridge voltage selecting switch 

This switch is to select the bridge supply voltage.   

You  can  make  your  selection  for  the  supply  voltage  among  DC  1V,  2V,  5V, 

7.5V, 10V. 

7

  Bridge voltage fine control volume 

This    controls establishing bridge voltage precisely. 

8

  Bridge voltage senscing selecting switch 

When  select  by  EXT  mode  as  switch  to  compensate  bridge    voltage  by 

Cable's  resistance  change,  bridge  voltage  by  Cable  length  change  moves 

automatically and is corrected. 

9

  Internal CAL Bridge resistance selecting switch 

This  switch  is  to  select  the  bridge  resistance  while  you  are  doing  bridge 

calibration. 

10

  CAL Strain value switch 

  It is a shunt switch to calibrate the bridge in two different types - 2000

µ

 and 

200

µ

strain. 

11

  GAIN selecting switch 

Summary of Contents for AM-1000

Page 1: ...YNAMIC STRAIN AMPLIFIER USER S MANUAL DACELL CO LTD Address 681 1 Cheoksan Ri Nami Myeon Cheongweon Gun Chung Buk 363 810 Korea TEL 82 43 260 2242 FAX 82 43 260 2245 http www dacell com E Mail dana da...

Page 2: ...11 4 1 POWER CONNECTION 11 4 2 SENSOR CONNECTION 11 4 3 INPUT HIGH LEVEL VOLTAGE 14 4 4 NOTES FOR SELECTING SENSOR CABLE 14 4 5 NOTES FOR DYNAMIC MEASUREMENT 14 4 6 NOTES FOR STATIC MEASUREMENT 15 4 7...

Page 3: ...to Zero Balance method is used for Zero Span This device can use connection for Quarter Half Full bridge type of strain gauge It is designed to install on standard 19 Sub rack 2 SPECIFICATION 1 Measur...

Page 4: ...alibration CAL Extent 1 F S 1 0mV V 2000 strain 0 1mV V 200 strain 13 Sensitivity adjustment GAIN X1 X10 X100 X400 X1000 within 1 at error each point SPAN X1 X11 continuous variableness FINE X1 X2 con...

Page 5: ...t within 0 05 F S Sensitivity change within 0 05 F S 20 MONITOR Digital Voltage Display Over 10V will be displayed in 10 00 increased or decreased in each 0 01 Less 10V will be displayed in 0 000 incr...

Page 6: ...7 Bridge voltage fine control volume 8 Bridge voltage senscing selecting switch 9 Internal CAL Bridge resistance selecting switch 10 CAL Strain value switch 11 GAIN selecting switch 12 GAIN control vo...

Page 7: ...2 Back Panel 19 Low current output load more than 2k 20 High current output load more than 200 21 Sensor input connector 22 Case earth terminal 23 Fuse 24 Power selecting switch 25 Power input connec...

Page 8: ...will appear on the LED and then zero will be set 6 Bridge voltage selecting switch This switch is to select the bridge supply voltage You can make your selection for the supply voltage among DC 1V 2V...

Page 9: ...ltage monitor terminal This can monitor Sensor output voltage 18 Signal output monitor terminal This can monitor output voltage of Amplifier 19 Low current output more than 2k load connector It is the...

Page 10: ...9 24 Power selecting switch It is a selection AC 110V 220V switch AC 220V when deliver from a warehouse 25 Power input connector This connector is to supply external AC power...

Page 11: ...ch and then supply power to the device If the lamp is not lighted please check the fuse and power selecting switch 4 2 Sensor connection Connector is MIL C MS3112 14 19 connector of MIL C 26482 Series...

Page 12: ...11 ST AM1000 INTERNAL CIRCUIT Internal Dummy External Dummy Quarter Bridge connection...

Page 13: ...12 Prefered Alternate Half Bridge connection 4 Gauge Transducer Full Bridge connection...

Page 14: ...rature change 3 The input stage of amplifier are SIG SIG Since their minimum input current is originated by this input impedance it should be over 50mA to carry out the measurement 4 4 Notes for selec...

Page 15: ...value can be changed by the surrounding temperature Please pay your special attention for this point Please carry out the following procedures 1 You must us 3 line for Quarter Bridge 2 To minimize the...

Page 16: ...ve gauge To control the sensor proper voltage of the amplifier please turn the bridge voltage switch 4 9 Bridge balance Generally when the bridge balance is too much out you cannot carry out the measu...

Page 17: ...ance or proper voltage while stress analysis is being done So shunt compensation is very important when you set the amplifier gain Also when you use the analogue transducer force torque and pressure e...

Page 18: ...17 L P F Characteristic GAIN Band Width Characteristic...

Reviews: