16.5. X-RAY TUBE
16.5.1. SPECIFICATIONS
Manufacturer
Canon
Model
D-054
Focus size
0.5
– IEC 60336
Equivalent filtration
0.003 ft Al equiv.
Anode angle
5º
Anode material
Tungsten
Maximum voltage
100 kVp
Thermal capacity
35 kJ
Max thermal capacity and cooling time
curve
See graphic thermal curve on item
Maximum current
24mA rectified by half or whole wave and
20mA constant potential
Maximum filament current
3,5A / 4,2V
Frequency
DC
Maximum exposure time
20s
Max anodic power
680kW
Maximum continuous thermal dissipation
2.2 kW
ATTENTION:
The X-ray tube Canon D-054 model , is the EXCLUSIVE USE of the EAGLE 3D CT
scanner
Note: If the temperature inside the tubehead reaches 55°C, exposure to X-rays will be
interrupted and a message will be displayed. The exposure will be enabled again after
the internal temperature of the generator reaches 54.5°C.