Initial calibration:

 On first use, the app will prompt to calibrate the pH probe, the 

Tank 

Water 

pump and the 

KH Solution 

pump. Follow the on-screen calibration instructions 

carefully. Please note this stage of the setup must be done accurately as failure to do 

so will cause inaccurate results. If the initial prompt does not automatically appear go to 

settings, select guide and follow the steps shown in the app.

•  

Once this initial calibration procedure is complete it is also possible to calibrate the 

Reagent 

pump. This is not essential but can achieve a greater accuracy in testing to 

perform this go to the settings page select ‘

Maintain

’ and then ‘

Calibration

’.

•  

Insert the stirring bead into the glass beaker and put the beaker in place in the  

KH Manager. It is important that the drain tube is positioned so that the end is touching 

the bottom of the beaker and that the water supply tube is located over the beaker.  

Re-attach the front cover plate and insert the probe. The pH probe will be held close to 

the front of the beaker and should be positioned so the tip sits just above the height of 

the stirring bead.

•  

In the 

Kamoer Remote

 app open the settings menu, go to ‘

Maintain

’ and then select 

Running Pumps

’, this allows you to run each pump head to confirm there is no air in the 

tubing. It is recommended to run the 

Tank Water 

pump (‘Saltwater Pump’) and

 Reagent 

pump to check that they are correctly dropping into the beaker and all air is purged. 

•  

The magnetic stirrer (‘Stirring Pump’) is factory set at 50% power. If when running, the 

stirring bead jumps and does not spin smoothly this setting should be reduced.

• 

 

Return to the main settings page, choose ‘

Test

’ from the bottom of the screen and select 

the    icon in the top right, this will allow you to set the number of times that the beaker 

will be flushed with tank water before each test, it is recommended to set this between 

3 and 4 times.

•  

Initial test:

 Return to the ‘

Manual

’ test page and press ‘

Start’

, this will initiate a one 

off manual test to confirm that the KH manager is operating correctly. It is highly 

recommended after any calibration and before setting the device for automatic testing 

to perform a manual test.

•  

The KH Manager is now ready to program for automatic monitoring and dosing.  

For detailed instructions on this and tips for use please view the KH Manager product 

page on our website 

www.theaquariumsolution.com 

or scan the QR code below.

Visit our website for tips  

and instructions for use

EN

Summary of Contents for KHManager

Page 1: ...FOLLOW US FOR INSPIRATION www theaquariumsolution com fb com theaquariumsolution theaquariumsolution...

Page 2: ...Q U I C K S T A R T G U I D E...

Page 3: ...LIMITED WARRANTY Warranty is only valid for units correctly installed and maintained all faults caused by improper use or water damage are not covered Should any defect in material or workmanship be f...

Page 4: ...e connected by Wi Fi to the cloud to allow remote connection or directly to your mobile device using Bluetooth for local control Remote Access Cloud Connection Ensure that your mobile device is connec...

Page 5: ...HE KH MANAGER MUST BE INSTALLED ON A FLAT LEVEL AND STABLE SURFACE Attach the tubing using the twist lock connectors as shown in the diagram above matching the coloured hose to the correct coloured po...

Page 6: ...In the Kamoer Remote app open the settings menu go to Maintain and then select Running Pumps this allows you to run each pump head to confirm there is no air in the tubing It is recommended to run the...

Page 7: ...lt werden verpflichtet sich D D The Aquarium Solution Ltd das defekte Teil kostenlos zu reparieren oder nach unserem Ermessen zu ersetzen Wir verfolgen eine Politik der st ndigen technischen Verbesser...

Page 8: ...n Sie den Anweisungen auf dem Bildschirm um die Registrierung und Anmeldung abzuschlie en DEN KH MANAGER VERBINDEN Der KH Manager kann ber Wi Fi mit der Cloud verbunden werden um eine Fernverbindung z...

Page 9: ...cannen Sobald sie gefunden werden werden sie auf dem Bildschirm angezeigt Tippen Sie auf das Symbol und die Seriennummer f r Ihren KH Manager Daraufhin werden Sie zu den Einstellungsseiten f r das Ger...

Page 10: ...deckplatte wieder an und setzen Sie die Sonde ein Die pH Sonde wird nahe an der Vorderseite des Bechers gehalten und sollte so positioniert werden dass die Spitze knapp ber der H he der R hrperle sitz...

Page 11: ...MIT E La garantie n est valable que pour les unit s correctement install es et entretenues Tous les d fauts caus s par une mauvaise utilisation ou un d g t des eaux ne sont pas couverts Si un d faut d...

Page 12: ...trement et la connexion CONNECTER LE KH MANAGER Le KH Manager peut tre connect au cloud via Wi Fi pour une connexion distance ou directement votre appareil mobile via Bluetooth pour un contr le local...

Page 13: ...ion de l appareil INSTALLATION DU KH MANAGER LE KH MANAGER DOIT TRE INSTALL SUR UNE SURFACE PLATE PLANE ET STABLE Fixez les tuyaux l aide des raccords Twist Lock comme indiqu sur l illustration ci des...

Page 14: ...ue pr s de l avant du b cher et doit tre positionn e de sorte que son extr mit se trouve juste au dessus du niveau de la perle d agitation Dans l application Kamoer Remote ouvrez le menu des param tre...

Page 15: ...www theaquariumsolution com Notes...

Page 16: ......

Reviews: