14
Pump head calibration
When the pump is initially set up, pump tubes have been replaced, or if the unit has been in
service for a period of time it is worth calibrating pump heads to ensure accuracy.
To calibrate open the app and select the device.
Tap the ‘Setting’ icon located at the bottom right of the page.
Select ‘Flow Calibration’ from the list and then choose the pump head to calibrate.
Make sure the inlet hose for the pump is submerged and that the outlet hose is directed into
a suitable collection container.
Tap on ‘Start emptying’ to prime the pump head and hose. Once all air has been expelled
and the line is fully primed tap ‘Emptying, click to stop’.
Move the outlet hose into a suitable measuring cylinder.
Set the time for pump calibration by selecting ‘Add time’ and scrolling through to the desired
time. This can be set up to 60 seconds. For increased accuracy, a longer calibration time is
best. Please be aware the measuring cylinder needs to be large enough to hold the volume
of the water for the calibration.
Tap on ‘Start add’ and the pump will run for the time selected in the previous step.