CytoTherm CT-DR Owner'S Manual Download Page 2

 

DESCRIPTION 

 

The CytoTherm DR (Model CT-DR) is a laboratory instrument for thawing plasma while keeping the plasma bag dry. 
Tempered water is recirculated through a bladder. The plasma bag to be thawed lays on one part of the bladder, while the 
rest of the bladder is folded over to cover the plasma bag. The plasma bag remains dry throughout. The rocking action 
speeds thawing. Plasma bag leakage is isolated to one of 3 separate sections. If a leak is detected, the thawing stops and an 
alarm sounds.  
 
A digital timer displays the time to completion of the thawing. The plasma bag is also visible during thawing so you can 
check when it has been fully thawed. 
 
Each of the 3 sections can thaw two 450ml. bags or one 1,000ml. Jumbo bag for a total of 6 x 450ml. or 3 Jumbo bags. 
 
All the air that comes in contact with the internal water passes through 0.1 micron AIR FILTER. 

 

SETTING-UP 

 

Make sure you have received the following: 
….. Control Panel 

 

 

….. Bath 

….. Rack 

 

 

 

….. 2 Arms 

….. Lid                                   

 

….. Sensor Connector  

….. 6 Bladders, 3 mounted inside RACK   ….. 4 x Air Filters 

 

….. Drain tubing with male drain fitting.   ….. Inlet and Outlet Port Plugs (3 of each) 
….. Power Cord                                           ….. Instruction Book  
….. Algicide (contiguous USA only)         
For instructional video, see www.cytotherm.com. 
 
Screw the ARMS to the side of the CONTROL SECTION. The notched  side of the ARMS face up.   
  
Place the BATH in front of the CONTROL SECTION so that the chassis on the BATH is next to the indented portion of 
the CONTROL SECTION. 
 
Connect the umbilical cord of the CONTROL SECTION to the plug receptacle on the back of the BATH section.  

* * THE POWER SWITCH MUST BE OFF 

WHEN CONNECTING OR DISCONNECTING THIS CONNECTION * * 

 
Lift the BATH and position it so that the pivots on the side of the BATH slip into the notches on the ARMS. 
 
Make sure the power switch is OFF. Connect the power cord to a properly grounded electrical outlet.  
 
 
Fill the bath with tap water  to 2.5Cm. (1 inch) above the pivot screw inside the BATH. If you use deionized water, add a 
pinch (0.01 gr) of salt to make the water conductive. Do NOT use saline solution.  Add 5 ml. of Algicide Part #ALG240 (1.0 
ml. per 4 liter of water available in contiguous USA only) to the water. Any other bath clearing agent can also be used to 
keep the water clear. 
 
Connect the pump output tubing to the input fitting on 
the bottom of the RACK. Place the RACK onto the top 
the BATH. The RACK has 3 bladders mounted inside. 
   
Connect one side of the SENSOR CONNECTOR to the 
CONTROL PANEL and the other end to the 2 rods on the 
left side of the RACK. The SENSOR CONNECTOR 
senses if the LID is closed and if there is a leak. 
 
Open the RACK by pushing the LOCK BAR to the left. 
Lift the LID by rotating it back, and when it is vertical, 
slowly lower it straight down into the slot.  

Summary of Contents for CT-DR

Page 1: ...well Ave Trenton NJ USA Tel 609 396 1456 Fax 609 396 9395 Do not forget to mail in your warranty card Please record the following Serial No ___________ Date of Purchase _________ TABLE OF CONTENTS Dra...

Page 2: ...ION The notched side of the ARMS face up Place the BATH in front of the CONTROL SECTION so that the chassis on the BATH is next to the indented portion of the CONTROL SECTION Connect the umbilical cor...

Page 3: ...hawed on the bottom part of the BLADDER Cover the plasma bag s with the top layer of the BLADDER Tuck the bladder in Close the LID by lifting straight up until it stops and then rotating it down forwa...

Page 4: ...leach solution Remount a new BLADDER You are ready to start thawing again Rinse off the bleach with water 3 Plasma bag and BLADDER both leaked It is a very unlikely case After you determine that the p...

Page 5: ...is 1 0C above the target temperature 5 Select the signal when thawing is done continuous or pulsed Enter PROGRAM MODE by holding ENTER while turning the unit ON Follow the displayed instructions to pr...

Page 6: ...Was the unit filled with water over 40C If yes then cool the water to 20C to reset the safety thermostat Check the Safety thermostat C TH and the heaters 14 77 for continuity Bag leaked but does not...

Page 7: ...nd displayed alarm that indicates that a plasma bag or BLADDER leaked MAINTENANCE Daily Wipe off the unit Every 3 Months Clean the BATH Unplug the unit Remove and discard the BLADDERS by first lifting...

Page 8: ...IBRATION Overtemperature Alarm Test Hold MINUS as you turn the unit ON DIAGNOSTICS will be displayed Push ENTER Push and release PLUS until ALARM TEST is displayed Push ENTER the heater will heat cont...

Page 9: ...water or add a pinch 0 01 gr of salt if you are using de ionized or distilled water OVER TEMPERATURE The unit will display TOO HOT and emit double beeps if the temperature reaches 1 0o C above the th...

Page 10: ...Power Cord Orn Blu Orn Blu Grn Yel Yel Mag Sensor Sensor Connector Level Sensor in Rack Level Sensor in Rack Start Stop Minus Plus Enter Reed Sensor Rock Position Thermistor Blu Grn Vio 2Grn 2Orn 2Br...