Parts and properties
Value
CIP
Max 50°C
Fluid line (tubing and wetted
components excl. tank)
Components and
piping
-0.4 bar g to 4.0 bar
1
Tank
-0.1 bar to 0.05 bar
1
Pump tubing PFD is limited to 2.5 bar.
System capacity
Data
Pump tubing PFB
Pump tubing PFD
Dimensions
3.2 mm ID × 6.4 mm OD
6.4 mm ID × 9.6 mm OD
Length
305 mm
305 mm
Filter inlet pressure
4.0 bar
1.5 bar
Feed flow range
20 to 150 mL/min
20 to 350 mL/min
Transfer flow range
1 to 50 mL/min
1
1 to 50 mL/min
1
Permeate flow range
1 to 50 mL/min
1
1 to 50 mL/min
1
TMP range produced
0.3 to 4.0 bar
0.3 to 2.5 bar
Hold-up volume
10 mL
17 mL
Minimum tank volume (tested)
4.4 mL
4.4 mL
Minimum working volume
15 mL
2
21 mL
1
To achieve required dynamic ranges, tube size changes may be required, if tube size changes are used, then pump
flow will be limited to a given range for a fitted tube size.
2
Minimum hold-up volume (excl. filter) with 2 × Tube 3/32" ID and pump tube 3.2 mm ID, 55 mm from feed pressure
sensor to filter and from filter to retentate pressure sensor, respectively 15 mL.
Longer tubing will be supplied with the instrument, and addition to the 55 mm length used in calculations above will
increase the volume accordingly.
Filter and CIP tubing
The following tubing is used for filtration or CIP.
Tag
AXM
AXH
MidGee
straight
MidGee
hoop
Kvick Start
cassette
Kvick Lab
Packet
CIP
MF11
x
MF22
x
MF32
x
x
11 Reference information
11.1 Specifications
ÄKTA flux s Operating Instructions 29046893 AG