WARNING
When using hazardous chemical and biological agents, take all
suitable protective measures, such as wearing protective glasses
and gloves resistant to the substances used. Follow local and/or
national regulations for safe operation and maintenance of
ÄKTA flux s.
WARNING
In ÄKTA flux s, prevent fluid leakage by tighten the Luer connection
firmly. Do not open the Luer connection by twisting, bending or
vibration. Do not open connectors when pressurized or during on-
going process.
WARNING
Use fume mask during tank opening when processing hazardous
liquids, to avoid possible exposure to aerosols.
CAUTION
Risk of pinching by moving parts inside the tank or inside the
pumps. Do not insert your fingers or other objects into the tank or
other moving parts while ÄKTA flux s is powered.
Flammable liquids and explosive
environment
WARNING
ÄKTA flux s is
not approved to handle flammable liquids.
WARNING
ÄKTA flux s is
not approved for work in a potentially explosive
atmosphere, in areas classified as Zone 0 to Zone 2 according to
IEC 60079-10 2002. ÄKTA flux s does not fulfill the requirements of
the ATEX Directive.
2 Safety instructions
2.1 Safety precautions
16
ÄKTA flux s Operating Instructions 29046893 AG