Illustration of Fraction collector F9-R
connector panel
1
2
2
Part
Function
1
Node ID switch
2
UniNet-9 communication and power supply
Connect the Fraction collector
Connect the UniNet cable between any of the
UniNet-9 ports on the back of the
instrument and on the back of the Fraction collector F9-R or Fraction collector F9-R, 2nd.
Make sure that all unused
UniNet-9 ports on the instrument are plugged.
Node ID
Node ID is a unit number designation that is used by the instrument to distinguish
between several units of the same type. All standard valves and available optional
modules are pre-configured to give the desired function. However, the function of a valve
or module can be changed by changing the Node ID.
Also, in a troubleshooting situation it may be useful to check the Node ID of the
respective valve or module.
Note:
The function of a valve or module is defined by a combination of the module
type and its Node ID, not by its physical position.
Check/change Node ID
The Node ID is set by positioning the arrow of the rotating switch at the back of the
Fraction collector. Use a screwdriver to set the arrow of the switch to the desired
number.
8
ÄKTA avant and ÄKTA pure Instructions 29021462 AD