Notes.
1.
Thicknesses stated are for steel with a velocity of 5920 m/s (0.2331 in/
μ
s), this
will vary with different metals/materials.
2.
S2C Probe using Deep Coat mode to measure through thick coatings.
3.
T5B & T2C Probes will include any coating thickness in the metal thickness
measurement when used in Single Echo mode.
Single Element Probes and Protective Membranes
All Cygnus single element probes have a soft face and are
therefore fitted with a Polyurethane Membrane which provides
better contact on rough surfaces and protects the probe face from
wear, prolonging the life of the probe.
Check the membrane regularly as it is important the membrane is
changed as soon as it shows any signs of wear.
Single Element Probe Membrane Parts.
Single Element Probe Membrane Locking Key
Probe Body
Locking Ring
Polyurethane Membrane
Knurled Ring