Shenzhen Zhiyong Electronics Co., Ltd.
11
10. Performance Verification
The below operation is for performance verification of the electric specification, requirement
for test equipment is shown below:
Figure 1 BNC-male-to-female-dual
Figure 2 BNC-female-to-dual binding post
Figure 3 BNC-male-to-dual binding post
Figure 4 BNC-male-to 50 Ω load
Equipment
Minimum Requirements
Usages
Oscilloscope
Bandwidth≥100MHZ; Accuracy
≦
1.5%, e.g.
Tektronix MSO/DSO4000
Displays probe output
Standard signal
generator; calibrator
Amplitude accuracy≤0.75%; rise time≤3ns
e.g.: FLUKE/WAVETEK 9100
Test bandwidth; AC
accuracy; common
mode rejection ration
Digital multimeter
Accuracy of not less than 6 and a half
e.g.
:
KEITHLEY 2000
Test the DC accuracy
Insulation pincer clips
Supplied in the accessories
Testing clips
BNC adapter 1
BNC-male-to-female-dual show as Figure 1
Test adapter
BNC adapter 2
BNC-male-to-dual binding post show as
Figure 2
Test adapter
BNC adapter 3
BNC-female-to-dual binding post show as
Figure 3
Test adapter
Load terminal
BNC-male-to 50Ωload show as Figure 4
Signal source load
CYBERTEK