![CyberResearch UMDAS 0802DA User Manual Download Page 20](http://html1.mh-extra.com/html/cyberresearch/umdas-0802da/umdas-0802da_user-manual_2703026020.webp)
In differential mode, the following two requirements must be met for linear operation:
!
Any analog input must remain in the
−
10V to +20V range with respect to ground at all times.
!
The maximum differential voltage on any given analog input pair must remain within the selected voltage
range.
The input [
common
-
mode
voltage
+
signal
] of the differential channel must be in the
−
10 V to +20 V range in order to yield a useful result. For example, you input a 4 V pp sine wave to CHHI, and
apply the same sine wave 180° out of phase to CHLO. The common mode voltage is 0 V. The differential input
voltage swings from 4 V
−
(
−
4 V) = 8 V to (
−
4 V)
−
4 V =
−
8V. Both inputs satisfy the
−
10 V to +20 V input range requirement, and the differential voltage is suited for the ±10 V input range (see
).
Figure 3-5. Differential voltage example: common mode voltage of 0 V
+4V
-4V
+4V
-4V
0V
CHLO
CHHI
+/-8V
8V Differential
Measured Signal
If you increase the common mode voltage to 11 V, the differential remains at ±8 V. Although the [
common
-
mode
voltage
+
signal
] on each input now has a range of +7 V to +15 V, both inputs still satisfy the
−
10 V to
+20 V input requirement (see
).
Figure 3-6. Differential voltage example: common mode voltage of 11 V
+11V
+7V
+11V
+/-8V
+15V
8V Differential
CHLO
CHHI
Measured Signal
If you decrease the common-mode voltage to
−
7 V, the differential stays at ±8 V. However, the solution now
violates the input range condition of
−
10 V to +20 V. The voltage on each analog input now swings from
−
3V
to
−
11V. Voltages between
−
10 V and
−
3 V are resolved, but those below -10 V are clipped (see
Figure 3-7. Differential voltage example: common mode voltage of -7 V
-3V
-11V
-7V
+/-7V
8V Differential
-11V
-7V
-3V
3V
CHLO
CHHI
Measured Signal
3-5
UMDAS 0802DA
User's Guid
e F
unctional Details