CyberMed M156S User’s Manual
v
Safety Instructions
Intended use
The CyberMed M156S is intended to serve as a medical monitor which is designed for general purpose for hospital environment. It
shall not be used for life-supporting system.
Intended User profile
The equipment is intended for infant or adults by profession Health care professionals.
Greeting & Setup
Thank you for purchasing the CyberMed M156S. We wish that this product will be durable and reliable in providing your medical
application needs. Please follow the instructions below to ensure the unit continues to have high performance.
Unpacking
After opening the carton, there will be a medical panel PC unit with an accessory box. Examine the contents to see if there are
damages to the unit and if all accessories are present.
Setting up
Please read this manual carefully and remember to keep this manual for future reference.
Safety Instructions & Cleaning
The unit has undergone various tests in order to comply with safety standards. Inappropriate use of the open frame unit may be
dangerous. Please remember to follow the instructions below to insure your safety during the installation and operating process.
Transporting & Placement of unit
1.
When moving the unit on a cart; be very cautious. Quick stops, excessive forces and uneven surfaces may cause the cart to
overturn thus risking the unit to fall to the ground.
2.
If the medical panel PC unit does fall to the ground, immediately turn the power off and disconnect cords. Then contact a service
technician for repairs. Continual use of the unit may result cause a fire or electric shock. Also, do not repair the unit on your
own.
3.
Having two or more people transporting the display unit is recommended. In addition, when installing the unit by suspending it
also requires two or more people.
4.
Before suspending the unit, make sure the material used for suspension is sturdy and stable. If not properly suspended, the
display unit may fall and cause serious injury to people standing nearby as well as to the unit itself.
5.
If you wish to mount the display unit, remember to use only the mounting hardware recommended by the manufacturer.
Electrical and Power Source Related
1.
This medical panel PC unit must operate on a power source as shown on the specification label. If you are not sure what type of
power supply used in the area, consult your dealer or local power supplier.
2.
The power cords must not be damaged. Applied pressure, added heat, and tugging may damage the power cord.
3.
The power cord must be routed properly when setup takes place. We advise that this aspect measure is to prevent people from
stepping on the cords or while the unit is suspended to prevent flying objects from getting tangled with the unit.
4.
For plug-in equipment, the power outlet socket must be located near the equipment and must be easily accessible.
5.
Do not overload the AC outlets or extension cords. Electrical shocks or fires may occur from overloading.
6.
Do not touch the power source during a thunderstorm.
7.
If your hands are wet, do not touch the plug.
8.
Use your thumb and index finger, grip firmly on the power cord to disconnect from the electrical socket. By pulling the power
cord, may result in damaging it.
9.
If the unit is not going to be in use for an extended period of time, remember to disconnect the unit.
10.
The medical panel PC unit uses voltage between 100-240VAC. Connect the unit to a power source with the same numerical
value as shown. Please use only the power cord provided by the dealer to ensure safety and EMC compliance.
Various Factors of Environment
1.
Do not insert objects into the openings.