PARTS REPLACEMENT PROCEDURES 

 

The  following  procedures  are  meant  to  aid  the  technician/engineer  in  replacing 
defective or damaged components.  These procedures are meant to be used by 
qualified  personnel  only.    Extreme  caution  should  be  used  and  all  necessary 
safety precautions taken when working on this equipment. 

 
 
 

CIRCUIT BREAKER REPLACEMENT 

 

1. 

Disconnect the illuminator from the power source. 

2. 

Remove the 12 cover screws on the sides of the unit. 

3. 

Using an ohmmeter verify which circuit breaker is defective. 

4. 

Remove the defective circuit breaker and replace with the new one. 

5. 

Replace the top cover.  

6. 

Replace the 12 cover screws and secure. 

 
 
 

SHUTTER REPLACEMENT 

 
1. 

Disconnect the illuminator from the power source. 

2. 

Remove the 12 cover screws in the sides of the unit.

   

3. 

Remove the 3 screws on the bottom of the unit which holds the front panel 
to the bottom housing.  

4. 

Remove the screw holding the shutter to the front panel. 

5. 

Pull off the shutter assembly. 

6. 

Replace the shutter belt if it shows signs of damage. 

7. 

Replace the shutter assembly with the new one. 

8. 

With  the  control  turned  fully  clockwise,  align  the  shutter  in  such  a  way  that  it 
does  not  cover  the  hole  in  the front  panel. Replace  the  screw  and  tighten  the 
screw until it holds the shutter in place and still allows for free movement of the 
shutter

. Do not over tighten the screw. 

9.        Replace the front panel and secure to the bottom housing with the 3 screws. 
10.      Replace the top cover.  
11.      Replace the 12 cover screws and secure. 
 
 
 

 
 

LIT054CUDA 

 

Rev. A

 

(English) 

 

Page 6 of 9 

Summary of Contents for MD-150T

Page 1: ...LIT054CUDA Rev A English MD 150T 150 Watt Dual Halogen Illuminator Service Manual 6018 Bowdendale Avenue Jacksonville FL 32116 Customer Service 904 737 7611 FAX 904 733 4832 Toll Free 877 677 2832...

Page 2: ...ICES 3 GENERAL THEORY OF OPERATION 4 INTERNAL VIEW 5 CIRCUIT BREAKER REPLACEMENT 6 SHUTTER REPLACEMENT 6 TRANSFORMER REPLACEMENT 7 COOLING FAN REPLACEMENT 7 ELECTRICAL SCHEMATIC Figure 1 8 REPLACEMENT...

Page 3: ...circuitry and wiring which could be hazardous with contact Use extreme caution when working on equipment that has power applied to it TECHNICAL SUPPORT SERVICES In the event that you experience diffi...

Page 4: ...circuit breakers A double pole interlock switch provides operator safety which is located on right side of the lamp heat shield INTENSITY CONTROL A rotating stainless steel disc that is placed in fro...

Page 5: ...LIT054CUDA Rev A English Page 5 of 9...

Page 6: ...1 Disconnect the illuminator from the power source 2 Remove the 12 cover screws in the sides of the unit 3 Remove the 3 screws on the bottom of the unit which holds the front panel to the bottom hous...

Page 7: ...sure the lamp access door is over the lamp cartridge 11 Replace the 12 cover screws and secure COOLING FAN REPLACEMENT 1 Disconnect the illuminator from the power source 2 Remove the 12 cover screws i...

Page 8: ...FIGURE 1 ELECTRICAL SCHEMATIC MD 150T DUAL HALOGEN ILLUMINATOR LIT054CUDA Rev A English Page 8 of 9...

Page 9: ...5 175 20 800 018 Circuit Breaker 5A 800 015 Shutter Assembly 800 003 Cooling Fan SP103A 800 019 Shutter belt Large 800 020 Shutter belt Small 800 021 For parts not listed here call our Technical Servi...

Reviews: