background image

 
 

LIT 232 

CUDA

®

 Surgical                                                                                                      Rev. -

 

 

(English)

                                                                                                              

Date of Revision:01/26/18

 

                                                                                                                                                     Page 5 of 6                                             

 

NOTE:  

  If  a  red  LED  is  lit  on  a  bay,  the  battery  of  the  corresponding  bay  should  be 

removed  from  the  charger  and  its  contacts  checked.  Install  the  battery  pack 
again.  If  the  red  LED  illuminates  again,  remove  the  battery  pack  from  the 
charger and contact Sunoptic Technologies for service.  

  The  unit  will  become  warm  during  normal  operation.  This  does  not  indicate  a 

malfunction.  

  Prevent metallic objects from coming into contact with the metal parts of the DC 

plug  of  the  power  supply. If this happens, a short may occur and the unit may 
be damaged.  

  Ensure  the  battery  has  proper  orientation  before  insertion.  Do  not  attempt  to 

force a battery into the slot as this may damage the contacts.  

 

5. 

CLEANING INSTRUCTIONS 

 

  Disconnect the Charger from the power source. 

  Clean  the  exterior  of  the  charger  with  a  cloth  that  has  been  dampened  (not 

dripping)  using  a  hospital  disinfectant.  Follow  applicable  hospital  protocols  during 
cleaning and dry immediately.

   

 

 

6. 

DISPOSAL/RECYCLE 

 

  In  accordance  with  the  European  Waste  from  Electrical  and  Electronic  Equipment 

(WEEE)  directive,  we  encourage  our  customers  to  recycle  this  product  whenever 
possible.  Disposal of this unit must be performed in accordance with the applicable 
local environmental regulations.  

  In the US a list of recyclers in your area can be found at: http:/www.eiae.org/. 

  Please  contact  our  Repair  Department  to  issue  a  return  authorization  to  return 

product to manufacturer at the end of product life. 

 

7. 

TROUBLESHOOTING 

SYMPTOM

 

POSSIBLE CAUSES/SOLUTIONS

 

Battery not 

charging

 

  Ensure charger power cord is plugged in to the AC power outlet. 

  Ensure DC connector is plugged into the back of the charger  

  Ensure battery is properly seated in the charger bay.  

  If  repeated  attempts  to  charge  the  battery  have  failed,  the  battery 

may be faulty. Replace the battery. 

  The  charger  may  have  a  faulty  power  cord  and/or  power  brick. 

Replace the power cord and/or power brick. 

Charger LED 

indicator not 

working

 

  Ensure charger power cord is plugged in to the AC power outlet. 

  Ensure battery is properly seated in the charger bay. 

  The  charger  may  have  a  faulty  power  cord  and/or  power  brick. 

Replace the power cord and/or power brick. 

 
 
 

Summary of Contents for LLS-3736-2

Page 1: ...1 of 6 LLS 3736 2 Two Bay Battery Charger Operator Manual ST Technologies 6018 Bowdendale Avenue Jacksonville FL 32216 USA Customer Service 904 208 2290 Toll Free 877 814 2237 EC REP RMS UK Ltd 28 Trinity Road Nailsea Somerset BS48 4NU United Kingdom TEL 01275 858891 ...

Page 2: ...sion 01 26 18 Page 2 of 6 TABLE OF CONTENTS 1 INTENDED USE INDICATIONS FOR USE 2 GENERAL WARNINGS 3 TECHNICAL SPECIFICATIONS 4 BATTERY CHARGING INSTRUCTIONS 5 CLEANING INSTRUCTIONS 6 DISPOSAL RECYCLE 7 TROUBLESHOOTING 8 LIMITED WARRANTY 9 SYMBOLOGY ...

Page 3: ...ected Do not allow water to collect in the bays or on top of the charger Do not allow water to enter the power connection on the back of the charger Keep clean and dry Do not short circuit crush open shred incinerate or dismantle the charger or the power supply brick The charger does not have serviceable parts Contact ST Technologies for replacement instructions Do not modify or change the power c...

Page 4: ...le surface Plug the output connector of the external AC DC power supply into the DC input located on the rear of the charging station Then plug the external AC DC power supply into an AC outlet The charge process will start after inserting a battery pack into the slot The charge indication will be displayed see Table 1 Figure 1 Indication Display LED Indicator Charging Condition None Off Mode Batt...

Page 5: ...ly 6 DISPOSAL RECYCLE In accordance with the European Waste from Electrical and Electronic Equipment WEEE directive we encourage our customers to recycle this product whenever possible Disposal of this unit must be performed in accordance with the applicable local environmental regulations In the US a list of recyclers in your area can be found at http www eiae org Please contact our Repair Depart...

Page 6: ...telephone number and a return address Warranty does not cover equipment subject to misuse accidental damage and normal wear and tear This warranty gives you specific legal rights and you may also have other rights that vary from state to state POST WARRANTY REPAIRS You may return your product s for repair shipping prepaid to the factory Your product will be inspected and an estimate of repair char...