Operation/Technical Manual
NORM-O-TEMP
®
, Model 111W
Page 11 of 82
NORM-O-TEMP
®
Operating Instructions Quick Start Guide
Read Operation Manual before using this device
First Time Use Cleaning:
Refer to section 5-1.1, 5-1.2 and 5-1.3 for cleaning instructions.
Initial Operation:
1. Check reservoir water level and fill if necessary
2. Insert power cord plug into hospital grade outlet
3. Connect hyperthermia blanket or pad
4. Turn power switch “ON”
Unit will go through an 8-10 second calibration cycle to test all indicators and will flash the
set point temperature of 37
°
C. Observe Water Flow Indicator on the side of the unit to verify
that the pump is circulating water when blanket/pad is connected. The unit will start heating
to a preset temperature of 37
°
C. To change the desired set point temperature, proceed to
step 5 below.
5. Depress the”TEMP SET” button. The display will flash the previous set point temperature
or the default temperature of 37
°
C if the unit has been turned OFF, and then back ON.
NOTE: You have five seconds to depress one of the arrows or the temperature readout will
return to the actual reservoir water temperature.
6. Depress the “UP”
▲
or “DOWN”
▼
arrow to raise or lower the set point temperature to the
desired setting.