Operation / Technical Manual
NORM-O-TEMP
®
, Model 111W
Page 40 of 82
5-1.
Maintenance of Water Reservoir
The dual compartment reservoir holds approximately 1½ gallons (5.7 liters) of distilled
water that remains in the unit between periods of use.
Once a month the water reservoir should be drained and replenished. The drain hose,
included in the packaging of the unit, is attached and the water is pumped out of the unit.
When the LO WATER alarm sounds, the remaining water is drained by gravity.
Distilled/sterile water (or distilled water-cleaning agent preparation) is then poured into
the reservoir.
Quarterly, at a minimum, the Internal Cleaning / Dry Storage Procedure, listed below,
should be conducted.
NOTE:
DO NOT USE DE-IONIZED WATER OR ALCOHOL.
Chemical Cleaning Circulation Chart
UNIT
MODE
TEMPERATURE
DURATION
NORM-O-TEMP
®
111W
Heating
38°C
5 Minutes
5-1.1.
Internal Clean using Bleach / Dry Storage Procedure
For circulating water units utilizing household bleach
.
The following procedure cleans the fluid circuit in these products. The required
tools/supplies are appropriate drain hose(s) and/or hose connector(s), household
bleach, distilled water, U.S.P. Grade Propylene Glycol, and the appropriate AC
electrical power
Procedure:
A.
Drain the water from the reservoir as described in Section 5-1.4.
B.
Turn unit off (“O” position).
C.
Fill the reservoir with the appropriate amount of household bleach per the
below chart.
CAUTION
For safe handling and use of chemicals follow manufacturer guidelines.
Always drain the NORM-O-TEMP
®
to a sanitary drain because bacteria may be present
in the unit’s water supply.
UNIT
BLEACH
WATER
NORM-O-TEMP
®
, 111W
168 Milliliters
5.4 Liters