background image

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

©

 2009 Cardiovascular Systems, Inc.  

 

 

 

 

 

90074-00.F 

Introduction

 

 
This document describes the components and use of the Diamondback 360º® Orbital Atherectomy Controller 
(Controller).  The Controller is a component of the Diamondback Orbital Atherectomy System (OAS).  Read this 
document thoroughly before using the OAS, Controller, or any other components.  Review the contraindications, 
restrictions, warnings, and precautions carefully.  Additional instructions and information are found in the 
Diamondback Orbital Atherectomy Device (OAD) Instructions for Use and package inserts included with separately 
packaged components.  Read all instructions completely before using the OAS or any components.   

 
Indications for Use 

The Diamondback Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as 
therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for 
percutaneous transluminal atherectomy.  
The OAS is also indicated for removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). 
 

Diamondback 360 Orbital Atherectomy System (OAS) Description  

 
CSI Supplied Equipment 

 

Orbital Atherectomy Device (OAD) 

 

Controller  

 

Guide Wire - as specified in the OAD Instructions for Use 

 

Gas input line and connector – connects from the controller to the air supply 

 
Additional Components 
These are independently obtained, as listed below.  Follow the original manufacturer’s instructions for each 
independently sourced component.    

 

Guide Catheter / Sheath Introducer – See the appropriate Instructions for Use for recommended guide catheter 
sheath introducer sizes. 

 

Compressed Gas Cylinder or house air supply – The gas must be air or nitrogen.   

 

Cylinder Pressure Regulator – The regulator must be capable of delivering at least 5 scfm (140 liter/min) at 85-
100 psi (5.8-6.9 bar).  

 

Normal Saline Solution for Injection – At least one full 1-liter bag must be available.                                                                             

 

Other equipment as needed for interventional procedures.   

 

 
Diamondback 360 Orbital Atherectomy Controller 

This document explains the function of the Orbital Atherectomy Controller (Controller).  Device specific instructions 
for use are available separately as packaged with the Orbital Atherectomy Device (OAD). 
 
The Orbital Atherectomy Controller (Controller) is an electro-mechanical control system.  The functions of the 
Controller include: 

 

Regulation of gas pressure to the turbine that controls Shaft/Crown rotational speed.. 

 

Recording of operation time including event time, and total time. 

 

Regulation of saline flow through the Sheath covering the Shaft. 

 
 

Contraindications, Restrictions, Warnings and Precautions 

 

Contraindications 

Refer to the OAD Instructions for Use for the associated contraindications. 
 

Restrictions 

 

The OAS should only be used by physicians who are experienced in peripheral angioplasty at his or her 
institution and trained on the use of the OAS. 

 

Caution: 

Federal law (USA) restricts this device to sale by or on the order of a physician. 

 
Warnings  

 

If mechanical failure of the Controller occurs before or during the atherectomy procedure, discontinue use 
immediately and return the product to Cardiovascular Systems, Inc.  Do not attempt to use a damaged OAD, 
Controller, or other system component.  Use of damaged components may result in system malfunction or 
patient injury. 

 

Summary of Contents for Diamondback DB-2000

Page 1: ... 2009 Cardiovascular Systems Inc 90074 00 F Controller Manual Part I Instructions for Use Model DB 2000 ...

Page 2: ...riate Instructions for Use for recommended guide catheter sheath introducer sizes Compressed Gas Cylinder or house air supply The gas must be air or nitrogen Cylinder Pressure Regulator The regulator must be capable of delivering at least 5 scfm 140 liter min at 85 100 psi 5 8 6 9 bar Normal Saline Solution for Injection At least one full 1 liter bag must be available Other equipment as needed for...

Page 3: ...oller box spill saline or other fluids on the Controller Saline or fluids inside the Controller may result in corrosion or electrical hazard Make sure the cylinder pressure regulator s fitting is compatible with the gas cylinder used Verify the type of gas coupling provided with the Controller prior to procuring a pressure regulator and connecting hose If necessary use suitable adapters to make th...

Page 4: ...m a user perspective about an active notification condition Therapy Status Area The Therapy Status Area displays the following therapy information Speed is the actual speed of the OAD in thousands of RPM Event time is the elapsed time between therapy OAD start and stop for the last therapy session When the turbine is started for the next session the event time is reset to zero Total time is the su...

Page 5: ... opens the compressed gas line and starts the turbine which results in Shaft Crown rotation Releasing the Foot Pedal closes the compressed gas line and stops Shaft Crown rotation The pump speed automatically resets to low pump speed within 7 seconds after release of the foot pedal Compressed Gas Cylinder and Regulator Additional equipment needed to operate the OAS includes a compressed gas cylinde...

Page 6: ...oller 2 Insert the other end of the power cord into a properly rated hospital grade receptacle Compressed Gas Cylinder Setup 1 Setup the compressed gas air or nitrogen regulator and tubing according to standard hospital protocol 2 Connect the gas tubing to the Gas Input connector on the back of the Controller Note Push down on the Release Button to open the connector before inserting the hose conn...

Page 7: ...using standard hospital procedure d Lift the roller pump lever to open the roller pump door e Place the saline tubing into the grooved tubing guides of the roller pump Ensure that the tubing guide located on top of the pump housing is appropriately adjusted for tubing diameter See figure A f Close the roller pump door See Figure B Figure A Figure B 5 Turn the Control Unit on by pressing the Power ...

Page 8: ...h a soft cloth dampened in water and mild detergent Isopropyl alcohol water mix 70 30 or Cydex should be used to clean and disinfect the Controller System from blood or other contaminants Do not immerse the Controller Monitor or Foot Pedal in fluid Avoid use of solvents or abrasive cleaners which may damage the Controller System components Controller Maintenance and Calibration The Controller does...

Page 9: ...roller has detected that output pressure exceeded maximum for 10 seconds Reselect lower speed or exchange device MAX SPEED EXCEEDED MAX SPEED EXCEEDED The controller has detected that output speed exceeded maximum for 10 seconds Reselect lower speed or exchange device NO SPEED DETECTED CHECK AND RETRY No Tachometer reading detected 1 Check fiber optic connections 2 Replace fiber optic SYSTEM FAILU...

Page 10: ...panying Documents On Off Equal Potential Terminal Protective Earth Electrical Safety Rating Type CF Applied Part IPX1 Fuse Manufacturer MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601 1 IEC EN 60601 1 CAN CSA C22 2 No 601 1 3MSP ...

Page 11: ...AGENTS AND VENDORS SHALL NOT BE LIABLE FOR ANY DIRECT INDIRECT INCIDENTAL OR CONSEQUENTIAL LOSS DAMAGE OR EXPENSE ARISING FROM OR RELATED TO THE USE OF THIS DEVICE No person has authority to bind Cardiovascular Systems Inc to any representation warranty or liability except as set forth in this Disclaimer of Warranty Cardiovascular Systems Inc may at its sole discretion replace any device that is d...

Page 12: ... 2009 Cardiovascular Systems Inc 90074 00 F Controller Manual Part II Service and Maintenance Model DB 2000 ...

Page 13: ...ree of protection against the ingress of water IPX1 Controller IPX0 Monitor IPX8 Footswitch Equipment not suitable for use in the presence of flammable anesthetics Operation Mode Continuous Storage Transport and Operating Conditions Storage and Transport 50 C to 60 C 58 F to 140 F 10 to 100 Relative Humidity 5000 to 1060 hPA Operating 10 to 40 C 50 to 104 F Electromagnetic Compliance Medical elect...

Page 14: ...roller has detected that output pressure exceeded maximum for 10 seconds Reselect lower speed or exchange device MAX SPEED EXCEEDED MAX SPEED EXCEEDED The controller has detected that output speed exceeded maximum for 10 seconds Reselect lower speed or exchange device NO SPEED DETECTED CHECK AND RETRY No Tachometer reading detected 1 Check fiber optic connections 2 Replace fiber optic SYSTEM FAILU...

Page 15: ...anying Documents On Off Equal Potential Terminal Protective Earth Electrical Safety Rating Type CF Applied Part IPX1 Fuse Manufacturer MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601 1 IEC EN 60601 1 CAN CSA C22 2 No 601 1 3MSP ...

Page 16: ...GENTS AND VENDORS SHALL NOT BE LIABLE FOR ANY DIRECT INDIRECT INCIDENTAL OR CONSEQUENTIAL LOSS DAMAGE OR EXPENSE ARISING FROM OR RELATED TO THE USE OF THIS DEVICE No person has authority to bind Cardiovascular Systems Inc to any representation warranty or liability except as set forth in this Disclaimer of Warranty Cardiovascular Systems Inc may at its sole discretion replace any device that is de...

Page 17: ...ation Statement 13 May 2008 As outlined in Cardiovascular Systems Inc Instructions for Use IFU supplied with each unit the Diamondback 360 DB 2000 controller does not require routine periodic maintenance or calibration For Questions Please Contact Director of Quality Assurance Cardiovascular Systems Inc St Paul Minnesota 55112 USA Telephone 651 259 1600 ...

Reviews: