IRGASON
®
Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
32
11.
Examine the measurements that were written down for span CO
2
, span
H
2
O, and zero air. Compute the drift in instrument gain using the
following equation:
meas
meas
actual
zero
span
span
gain
−
=
where,
•
span
actual
= known concentration of the span gas
•
span
meas
= measured concentration of the span gas
•
zero
meas
= measured concentration in zero gas.
Note that in the zero-and- span window of
ECMon
, span
actual
is reported to
the right of the box where the user enters the span dew-point temperature.
The software calculates span
actual
by taking into account the dew-point
temperature and current ambient temperature and pressure. The equations
used for this calculation may be found in Appendix
. If drift (offset or
gain) for CO
2
or H
2
O is excessive, it may be time to replace the desiccant
and CO
2
scrubber bottles (see Section
(p. 32)
).
12.
With zero air still flowing and measurements stabilized, click on the
Zero
CO
2
and H
2
O
button in the
ECMon
zero-and-span window.
Air flow into the shroud should be close to the recommended rate.
If the flow is too low, the shroud will not be properly flushed; if it
is too high, the air pressure within the shroud will be too high, and
the analyzer will not be zeroed and spanned properly.
13.
Remove the zero air source and replace it with the CO
2
span gas. Allow
the gas to flow through the shroud, maintaining a flow between 0.4 and 0.6
l/min. Wait for readings to stabilize.
14.
In the zero-and-span window, enter the known concentration of CO
2
(in
ppm) in the box labeled
Span Concentration (dry)
and press
Span
.
15.
Replace the CO
2
span gas with an H
2
O span gas of known dew point.
Allow the gas to flow through the shroud; higher flows may be desired for
a couple of minutes to more quickly establish equilibrium before resuming
a flow between 0.4 and 0.6 l/min. Wait for the readings to stabilize.
16.
Enter the known dew point (in °C) in the box labeled
Span Dew Point
and
press
Span
.
17.
The zero-and-span procedure is now complete. Remove the shroud,
reconnect the IRGASON
®
temperature probe, and prepare the site for
normal operation.
9.5 Replacing Desiccant and Scrubber Bottles
If more than two years have passed since replacing the desiccant / scrubber
bottles within the IRGASON, or if the zero-and-span readings have drifted
NOTE
Summary of Contents for IRGASON
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 12: ...Table of Contents iv ...
Page 60: ...Appendix A Filter Bandwidth and Time Delay A 4 ...
Page 64: ...Appendix C Material Safety Data Sheets MSDS C 2 ...
Page 65: ...Appendix C Material Safety Data Sheets MSDS C 3 ...
Page 66: ...Appendix C Material Safety Data Sheets MSDS C 4 ...
Page 67: ...Appendix C Material Safety Data Sheets MSDS C 5 ...
Page 68: ...Appendix C Material Safety Data Sheets MSDS C 6 ...
Page 69: ...Appendix C Material Safety Data Sheets MSDS C 7 ...
Page 70: ...Appendix C Material Safety Data Sheets MSDS C 8 C 2 Decarbite MSDS ...
Page 71: ...Appendix C Material Safety Data Sheets MSDS C 9 ...
Page 72: ...Appendix C Material Safety Data Sheets MSDS C 10 ...
Page 73: ......