17
any integer value between 0 and 14. The value 15 is reserved as an SDM-group
trigger.
7.4 Unprompted Output
If the EC100 is to output data as USB or RS-485 (see Section
(p. 22)
), this setting must be set to
USB Port
or
RS485
, respectively.
The factory default is to disable the unprompted output and assume data are to
be logged via SDM (see Section
(p. 22)
Only one unprompted-output type (i.e., USB or RS-485) may be selected at a
given time. The rate at which the EC100 outputs these data is determined by
the
Unprompted Output Rate
setting.
7.5 Unprompted Output Rate
This setting determines the output rate for unprompted output (USB or RS-485;
(p. 22)
). If the unprompted output is
disabled, this parameter is not used. The factory default output rate is 10 Hz,
but it may be set to 10, 20, or 60 Hz.
7.6 RS-485 Baud Rate
If the unprompted output mode is set to RS-485, this parameter determines the
baud rate. Otherwise, this setting is not used. The RS-485 baud rate defaults to
115200 bps, although the user may enter another value.
7.7 Analog Output
The EC100 has two analog outputs for CO
2
and H
2
(p. 24)
, for more information). These outputs may be enabled or
disabled with this setting. The default is for analog output to be disabled. Note
that the analog output does not include sonic data.
7.8 ECMon Update Rate
This setting determines the rate at which data are sent over the USB connection
to the PC while running
ECMon
. The default setting of 10 Hz should be
adequate in most situations.
7.9 Temperature Sensor
This setting configures the EC100 electronics to integrate with the IRGASON.
The IRGASON measures ambient air temperature using a thermistor
temperature sensor. With
Auto-Select
selected, the EC100 will detect that the
IRGASON is connected and will report ambient air temperature measurements
from the thermistor.
To diagnose problems with the temperature measurement, a fixed temperature
value may be used, or the temperature sensor may be selected manually.
7.10 Fixed Temperature Value
If the
Temperature Sensor
setting is
None
, the IRGASON will use the value
of this setting for the sample temperature. This mode is intended for
troubleshooting only. In normal operation,
Temperature Sensor
is set to
Auto-Select
, and this setting is not used.
Summary of Contents for IRGASON
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 12: ...Table of Contents iv ...
Page 60: ...Appendix A Filter Bandwidth and Time Delay A 4 ...
Page 64: ...Appendix C Material Safety Data Sheets MSDS C 2 ...
Page 65: ...Appendix C Material Safety Data Sheets MSDS C 3 ...
Page 66: ...Appendix C Material Safety Data Sheets MSDS C 4 ...
Page 67: ...Appendix C Material Safety Data Sheets MSDS C 5 ...
Page 68: ...Appendix C Material Safety Data Sheets MSDS C 6 ...
Page 69: ...Appendix C Material Safety Data Sheets MSDS C 7 ...
Page 70: ...Appendix C Material Safety Data Sheets MSDS C 8 C 2 Decarbite MSDS ...
Page 71: ...Appendix C Material Safety Data Sheets MSDS C 9 ...
Page 72: ...Appendix C Material Safety Data Sheets MSDS C 10 ...
Page 73: ......