IRGASON
®
Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
22
8. EC100 Outputs
The EC100 is the IRGASON electronics module that measures the sensors and
controls the system. It outputs data in one of four types: SDM, USB, RS-485,
or analog. In general, Campbell Scientific recommends that SDM be used if a
Campbell Scientific datalogger is responsible for data collection. However,
RS-485 output is recommended if cable lengths exceed 100 meters. If a PC is
being used as the collection vehicle, USB and RS-485 are suitable outputs.
Analog output may also be used; however, only CO
2
density and H
2
O density
will be output. Information regarding each output type is provided in the
following sections.
8.1 SDM Output
SDM (Synchronous Device for Measurement) is a Campbell Scientific
communication protocol that allows synchronized measurement and rapid
communication between a Campbell Scientific datalogger and multiple devices
including the IRGASON. Although nearly all of the Campbell Scientific
dataloggers support SDM, only the CR6, CR1000, CR3000, and CR5000
dataloggers support the IRGASON.
To use SDM data output, connect an SDM cable from the EC100 (see Section
(p. 11)
) to a CR6, CR1000, CR3000, or CR5000
datalogger. On CR1000 dataloggers, the SDM protocol uses ports C1, C2, and
C3. These are multipurpose control ports that are SDM-activated when an
SDM instruction is used in the datalogger program. On CR3000 and CR5000
dataloggers, the SDM protocol uses SDM-dedicated ports SDM-C1, SDM-C2,
and SDM-C3. On CR6 dataloggers, the user must assign which set of three
ports will be used for SDM. This is done in the datalogger program using the
SDMBeginPort()
programming instruction. See the CR6 manual for more
details.
Each SDM device on the SDM bus must have a unique address. The
IRGASON has a factory default SDM address of 1, but may be changed to any
integer value between 0 and 14 (see Section
(p. 16)
).
The sample rate for SDM output is determined by the inverse of the datalogger
scan interval, as set by the user in the datalogger program. Data are output from
the EC100 when a request is received from the logger, i.e. a
prompted
output
mode. The number of data values sent from the EC100 to the datalogger is also
set by the user in the datalogger program. CRBasic, the programming language
used by Campbell Scientific dataloggers, uses the
EC100()
instruction to get
data from an IRGASON. This instruction is explained in detail under Section
(p. 34)
, of this manual.
8.2 USB or RS-485 Output
In contrast to the SDM output mode, which is prompted by a datalogger, data
can also be output from the EC100 via USB or RS-485 in an unprompted
mode. In this case, the EC100 sends out data without initiation from the
receiving device at a rate determined by the EC100. Only one unprompted
output type, USB or RS-485, may be selected at a given time. RS-485 output is
recommended if cable lengths exceed 100 meters. If a Campbell Scientific
datalogger is not being used to collect the data from the IRGASON, either
unprompted mode is recommended.
Summary of Contents for IRGASON
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 12: ...Table of Contents iv ...
Page 60: ...Appendix A Filter Bandwidth and Time Delay A 4 ...
Page 64: ...Appendix C Material Safety Data Sheets MSDS C 2 ...
Page 65: ...Appendix C Material Safety Data Sheets MSDS C 3 ...
Page 66: ...Appendix C Material Safety Data Sheets MSDS C 4 ...
Page 67: ...Appendix C Material Safety Data Sheets MSDS C 5 ...
Page 68: ...Appendix C Material Safety Data Sheets MSDS C 6 ...
Page 69: ...Appendix C Material Safety Data Sheets MSDS C 7 ...
Page 70: ...Appendix C Material Safety Data Sheets MSDS C 8 C 2 Decarbite MSDS ...
Page 71: ...Appendix C Material Safety Data Sheets MSDS C 9 ...
Page 72: ...Appendix C Material Safety Data Sheets MSDS C 10 ...
Page 73: ......