Operation
VX 570 English V1.10
Seite 30 von 44
11.3.1.4 Extended Sensor Settings
11.3.1.4.1 Definition of the reference settings
Here can be defined the desired measured media reference conditions for pressure,
temperature and times for the filter and averaging.
Setup
→
Sensor Setup
→
Advanced
Remark:
•
Factory pre-setting for reference temperature and reference pressure are 20 °C, 1000
hPa
•
All volume flow values (m³/h) and consumption values indicated in the display are
related to 20 °C and 1000 hPa (according to ISO 1217 intake condition)
•
Alternatively 0 °C and 1013 hPa (=standard cubic meter) can also be entered as a
reference.
•
Do not enter the operation pressure or the operation temperature under reference
conditions
!
Setup
→
Sensor Setup
→
Advanced
→
Ref. Settings
To make changes, first select a menu item with
the
„>“
key and then select it with the
„OK
key.
With
„set to Default“
the reference conditions
are set back to
1000mbar / 20°C.