A Quick Start Guide to the Model 18i / 14i / 12i
Cryogenic Temperature Monitor
USER DOCUMENTATION AND UTILITY SOFTWARE
User documentation, utility software, LabView drivers and sensor curves
are on the CD provided. Additionally, they are available at:
http://www.cryocon.com/CCdownload/CustomerCD12.zip
BASIC FRONT PANEL OPERATION
Pressing the
Power
(
) key for two seconds will toggle the monitor's
power on and off.
Several formats of the home screen are available to display different
information. Scroll through these formats by pressing the
Right
(
►
) key.
Setup and configuration functions are performed by working with the
monitor’s menu tree. To access this tree from the Home display, press
the
Enter
(
) key.
Navigation through all menus is performed by pressing the
Inc
(
▲
)
or
Dec
(
▼
) keys. A cursor will scroll up or down to show additional lines.
Moving up the tree is done by pressing the
Right
(
►
) key. The Home
display is at the top of the tree.
The left most character on each line of a menu is the cursor. These
indicate as follows:
Pressing the
Enter
(
) key will select the function described on the
menu line.
+
Pressing the
Enter
(
) key will cause the cursor to flash. Then,
pressing the
Inc
(
▲
)
or
DEC (▼)
keys will sequence through the
allowed choices for the line. To make a selection, press the
Enter
(
)
key again. To abort the selection process, press the
Right
(
►
) key.
Indicates a numeric entry field. To change the value displayed,
press the
Enter
(
) key and the cursor will flash. Then, press he
Inc
(
▲
) key to increment the number or the
Dec
(
▼
)
key to decrement
the number. When the desired value is shown, press the
Enter
(
)
key.
Key
Description
Enter
1) From Home screen, go to the top level setup menu.
2) Within a setup menu, enter data or select a field.
3) Press and hold this key for two seconds to toggle power.
▲
INC
1) Scroll display up. 2) When in a field selection mode, abort
entry and return to scroll mode. 3) In a numeric field,
increment.
▼
DEC
1) Scroll display down. 2) When in a field selection mode,
abort entry and return to scroll mode. 3) In a numeric field,
decrement.
►
Right
1) Move up the menu tree one level. 2) In selection mode,
scroll to next selection.
TEMPERATURE DISPLAYS
A sensor fault condition is identified by a display of
a clipping or clamping message. This usually
indicates that the sensor is open, disconnected or
shorted.
If a temperature reading is within the measurement
range of the instrument but is not within the
specified Sensor Calibration Curve, a display of
seven dot (.) characters is shown.
LED INDICATORS
There are three LED indicators below the display. They indicate the
following:
Alarm
(Red) – An enabled alarm condition is asserted.
Relay 1
(Green) and
Relay 2
(Green) - Relay asserted.
Latched alarms and relays are cleared by pressing the
Enter
(
) key.
ETHERNET COMMUNICATION
An Ethernet connection can be established by using a standard RJ-45
patch cable. The monitor acts as a server with a default IP and subnet
mask is
192.168.1.4
and
255.255.255.0
. The user can change this from
the front panel via a menu or from a web browser. DHCP can also be
selected.
To view the embedded web page, open a web browser and type the
default IP of 192.168.1.4 in the address bar.
The default port for the TCP server is 5000. The UDP server is at TCP+1.
WEB SERVER FUNCTIONS
Note:
The embedded web server's default user name is
admin
and default password is
cryocon
.
This secure server provides most of the functions required to view and
configure the instrument including the downloading of user-defined
sensors, real-time strip-charts, data-logging and firmware updates.
Clip
!!CLAMP!!